Fluke 5720A service manual Press the “DC V” softkey to bring up the next menu shown below

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 221
Image 221
4. Press the “DC V” softkey to bring up the next menu shown below:

Calibration and Verification 3

Range Calibration

PREV

MENU

IF3-2.EPS

3. Press the “Range Adjust” softkey to bring up the next menu shown below:

PREV

MENU

IF3-18.EPS

4. Press the “DC V” softkey to bring up the next menu shown below:

PREV

MENU

IF3-19.EPS

5.Press the “NEXT Menu” softkey, which scrolls through all the available dc voltage ranges, until 220V appears. Then press that selection’s softkey to open a display similar to the following:

PREV

MENU

IF3-20.EPS

6.Connect the 732B, null detector, and 752A in a 10:1 configuration, as shown in Figure 3-5.

7.Multiply the 732B’s value by 10, and enter this new value. (This value is the output of the 752A, to which you will null the calibrator’s output.) Then press Eto bring up the following display.

3-15

Page 221
Image 221
Fluke 5720A service manual Press the “DC V” softkey to bring up the next menu shown below