3-9. Calibration Procedure

5700A/5720A Series II Calibrator

Service Manual

3-9. Calibration Procedure

Before you being this procedure make sure the calibrator is powered on and has completed the appropriate warm-up period.

After you have finished calibration, but before you save the new constants, the calibrator presents the new changes as ± ppm, and as a percentage of specifications for each range and function. A list of changes can be sent to a computer through either the serial or instrument control (IEEE-488) port, or printed through the serial port. The largest proposed change will be displayed on the calibrator’s front panel.

Follow this procedure to calibrate the main output functions.

1.Press the “Setup Menus” softkey; then press the “Cal” softkey. The calibration menu appears as shown below:

PREV

MENU

IF3-1.EPS

2. Press the “Cal” softkey. The display shows the following.

PREV

MENU

3.To calibrate the main output functions, press one of the softkeys under “Calibration.” The display shows the following:

PREV

MENU

IF3-3.EPS