Fluke 5720A Calibration and Verification, Determining Test Limits for Other Calibration Intervals

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 273
Image 273
Table 3-32. Wideband Flatness Test Record (cont)

Calibration and Verification 3

Determining Test Limits for Other Calibration Intervals

Table 3-32. Wideband Flatness Test Record (cont)

Output

Level

30 mV

10 mV

3 mV

1 mV

Frequency

10kHz

120kHz

500kHz

2MHz

5MHz

10MHz

20MHz

30MHz

10kHz

120kHz

500kHz

2MHz

5MHz

10MHz

20MHz

30MHz

10kHz

120kHz

500kHz

2MHz

5MHz

10MHZ

20MHz

30MHz

10kHz

120kHz

500kHz

2MHz

5MHZ

10MHz

20MHz

30MHz

Measured

Flatness Error

Spec

±ppm

(90-day)

1000

1000

1100

1100

2100

2100

4100

10100

1000

1000

1300

1300

2300

2300

4300

10300

1000

1000

2000

2000

4000

4000

6000

16000

1000

1000

5000

5000

7000

7000

9000

30000

5790A

1-Year

Spec

500

500

500

500

1000

1000

1500

3500

500

500

700

700

1000

1000

1700

3700

500

500

1033

1033

1333

1333

2033

4033

500

500

1700

1700

2700

2700

4000

9000

TUR

2.00

2.00

2.20

2.20

2.10

2.10

2.73

2.89

2.00

2.00

1.86

1.86

2.30

2.30

2.53

2.78

2.00

2.00

1.94

1.94

3.00

3.00

2.95

3.97

2.00

2.00

2.94

2.94

2.59

2.59

2.25

3.33

GB

Factor

0.932

0.932

0.945

0.945

0.939

0.939

0.972

0.977

0.932

0.932

0.920

0.920

0.950

0.950

0.962

0.972

0.932

0.932

0.928

0.928

0.980

0.980

0.978

0.999

0.932

0.932

0.978

0.978

0.965

0.965

0.948

0.988

Test Limit

±ppm

932

932

1040

1040

1972

1972

3985

9868

932

932

1196

1196

2185

2185

4137

10012

932

932

1856

1856

3920

3920

5868

15984

932

932

4890

4890

6755

6755

8532

29640

3-67

Page 273
Image 273
Fluke 5720A Calibration and Verification, Determining Test Limits for Other Calibration Intervals, Output Level, Frequency