Fluke 5720A Calibration and Verification, Determining Test Limits for Other Calibration Intervals

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 271
Image 271
Table 3-30. Wideband Frequency Accuracy Test Record

Calibration and Verification 3

Determining Test Limits for Other Calibration Intervals

Table 3-30. Wideband Frequency Accuracy Test Record

Frequency ( Hz )

10Hz

100Hz

300Hz

500Hz

800Hz

900Hz

1kHz 1.19 kHz 2.2 MHz 3.5 MHz 3.8 MHz

10MHz

20MHz

30MHz

Frequency Measured

Tolerance Limits

99.99ms to 100.01 ms

9.999 ms to 10.001 ms

299.97Hz to 300.03 Hz

499.95Hz to 500.05 Hz

799.92Hz to 800.08 Hz

899.91Hz to 900.09 Hz

999.0Hz to 1.0001 kHz 1.189881 kHz to 1.190119 kHz 2.19978 MHz to 2.20022 MHz 3.49965 MHz to 3.50035 MHz 3.79962 MHz to 3.80038 MHz 9.990 MHz to 10.001 MHz 19.998 MHz to 20.002 MHz 29.997 MHz to 30.003 MHz

3-65

Page 271
Image 271
Fluke 5720A Calibration and Verification, Determining Test Limits for Other Calibration Intervals, Frequency Hz