5700A/5720A Series II Calibrator

Service Manual

approximately a 50% duty cycle. Connect the oscilloscope to pin 6 of U14 and verify it to be the compliment signal. If either signal is incorrect, U14 is probably at fault.

1 V

Figure 5-22. Waveform at Pin 10 of U6 10 ms Manual background

F5-22.EPS

Figure 5-22. Waveform at Pin 10 of U6

9.Check CH1 SHUNT. Connect an oscilloscope to TP6 (common to TP3). Set the Calibrator to 6.5V dc, operate, and set the oscilloscope to 2V/div at 2 ms/div. The oscilloscope should display a signal similar to Figure 5-25. If a failure is detected, check Q35 and its associated components.