Fluke 5720A Wideband Frequency Accuracy Test, Wideband AC Voltage Module Output Verification

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 244
Image 244
3-26. Wideband Frequency Accuracy Test

5700A/5720A Series II Calibrator

Service Manual

Note

The input impedance at INPUT 2 of the 5790A on the millivolt ranges is 10 MΩ shunted by 83 pF and the cable used to connect the shunt resistor to the 5790A has 40 pF capacitance, for a total of 123 pF.

The reactance of 123 pF at 10 kHz is 129 kΩ, and using formula (1), in the case where R=2 kΩ, the network impedance Z = 1.999760 kΩ. This produces an error of 120 PPM. The allowable error at 10 kHz is reduced to account for this error.

Using formula (2) we get a result of Xc = 88 kΩ, and a network impedance of Z =

1.9995 kΩ. This produces an error of approximately 250 ppm, or 0.025%. However, if we are making a measurement of 0.21% uncertainty (as in the present case), the ratio of measurement uncertainty is about 10:1 and is not of concern. It is easy to improve measurement uncertainty if desired, though, by measuring the actual input capacitance of the detector and any stray capacitance from input leads, etc. and making corrections. In this discussion, the UUT is assumed to be a high-impedance current source (like the 5700A), which can easily be verified by the manufacturer’s specifications, i.e., a specified voltage adder for current outputs above a given burden voltage.

3-26. Wideband Frequency Accuracy Test

Use Table 3-30 for the test record. Proceed as follows to test the Wideband module frequency accuracy:

1.Connect the Philips PM 6669 to the Calibrator wideband output and measure the output frequency at the frequencies listed in Table 3-30.

2.Verify that the frequency counter indicates frequencies within the 0.01% limits shown.

3-27. Wideband AC Voltage Module Output Verification

The wideband tests are for units with the Option 5700A-03 Wideband AC Module only. The verification test for the wideband module works as follows:

Accuracy at 1 kHz: Output at 1 kHz is tested by comparing the wideband output at the end of the cable and termination supplied with the instrument to the 5790A at INPUT 2.

Attenuator flatness: The attenuator flatness is tested using the 5790A wideband input and using reduced spec limits when the TUR (Test Uncertainty Ratio) is less than 4:1.

Table 3-11 lists the equipment required for testing and calibrating the Wideband module.

3-38

Page 244
Image 244
Fluke 5720A service manual Wideband Frequency Accuracy Test, Wideband AC Voltage Module Output Verification