Table 3-28. 5720A AC Current 2 mA and 200 μA Accuracy Test Record

5700A/5720A Series II Calibrator

Service Manual

Table 3-28. 5720A AC Current 2 mA and 200 μA Accuracy Test Record

Output Current

2mA

2mA

2mA

2mA

2mA

2mA

200μA

200μA

200μA

200μA

200μA

200μA

Frequency

10Hz

20Hz

40Hz

1kHz

5kHz

10kHz

10Hz

20Hz

40Hz

1kHz

5kHz

10kHz

Error

Spec.

±ppm

(90-day)

305

200

150

150

285

1500

380

240

180

180

395

1500

5790A

Transfer

Spec

±ppm

210

73

27

27

27

27

210

73

27

27

27

27

TUR

1.45

2.74

5.56

5.56

10.56

55.56

1.81

3.29

6.67

6.67

14.63

55.56

GB

Factor

0.874

0.971

1.000

1.000

1.000

1.000

0.916

0.988

1.000

1.000

1.000

1.000

Test Limits

±ppm

267

194

150

150

285

1500

348

237

180

180

+365/-425 (1) +1380/-1620

(1)

1:The spec is modified because of the capacitive loading of the 2K resistor by the cable and input of the 5790A.

Table 3-29. 5700A AC Current 2 mA and 200 μA Accuracy Test Record

Output Current

2mA

2mA

2mA

2mA

2mA

2mA

200μA

200μA

200μA

200μA

200μA

200μA

Frequency

10Hz

20Hz

40Hz

1kHz

5kHz

10kHz

10Hz

20Hz

40Hz

1kHz

5kHz

10kHz

Error

Spec

±ppm

(90-day)

725

400

160

160

850

2100

850

505

240

240

850

2100

5790A

Transfer

Spec

±ppm

210

73

27

27

27

27

210

73

27

27

27

27

TUR

3.45

5.48

5.93

5.93

31.48

77.8

4.05

6.92

8.89

8.89

31.48

77.8

GB

Factor

0.991

1.000

1.000

1.000

1.000

1.000

1.000

1.000

1.000

1.000

1.000

1.000

Test Limits

±ppm

718

400

160

160

850

2100

850

505

240

240

+820/-880 (1) +1980/-2220

(1)

1:The spec is modified because of the capacitive loading of the 2K resistor by the cable and input of the 5790A.

3-64

Page 270
Image 270
Fluke 5700A/5720A Series II Calibrator, 28. 5720A AC Current 2 mA and 200 μA Accuracy Test Record, Output Current