Fluke 5720A Calibration and Verification, Determining Test Limits for Other Calibration Intervals

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 275
Image 275
Table 3-34. Load Regulation Test Record

Calibration and Verification 3

Determining Test Limits for Other Calibration Intervals

Table 3-34. Load Regulation Test Record

Div. Setting

10V

100V

1000V

5700A/5720A Series II

Range

11V

220V

1100V

5700A/5720A Series II

Out/Full Load

10V/199Ω

100V/1999Ω

1000V/49.99 kΩ

Change In Null Measured

Change In Null Limit

±3.1 μV

±42 μV

±310 μV

Table 3-35. DC Voltage Linearity Test Record

Kelvin-Varley Setting

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

5700A/5720A Series II

Output

1V

2V

3V

4V

5V

6V

7V

8V

9V

Null Detector

Reading

Null Detector Reading limit

±2.3 μV

±2.6 μV

±2.9 μV

±3.2 μV

±3.5 μV

±3.8 μV

±4.1 μV

±4.4 μV

±4.7 μV

Table 3-36. DC Voltage Output Noise Test

Differential Amplifier

Sensitivity

50μV/division

50μV/division

50μV/division

100μV/division

500μV/division

Calibrator (UUT)

2.2V

10V

20V

200V

1000V

Rms Meter

Reading

Maximum rms Meter

Reading

150mV

500mV

500mV

750mV

500mV

3-69

Page 275
Image 275
Fluke 5720A Calibration and Verification, Determining Test Limits for Other Calibration Intervals, Div. Setting