Fluke 5720A service manual 2. Waveform at TP1, F5-2.EPS

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 335
Image 335
F5-2.EPS

Troubleshooting 5

Component-level Troubleshooting

logic high by measuring U15 pin 5. If the signal is incorrect, the control line WB ON/OFF* from the Wideband Output (A5) assembly is probably at fault. If the control line is correct check U15A and associated components.

Note

The input CLK and CLK* is a low-level (400 mV p-p) signal generated by the Regulator/Guard Crossing assembly.

4.Check the 4 MHz to 8 MHz feedback signal. Set the Calibrator to 1V at 1.2 MHz, operate. Connect an oscilloscope to TP15 and verify it displays a 4.8 MHz signal similar to that shown in Figure 5-3. Use a frequency counter and verify that the frequency of this signal is 4.8 MHz ±0.01%. If a failure is detected, continue with step 5; otherwise skip to step 10.

1 V

Figure 5-2. Waveform at TP1 10 ms Manual background

F5-2.EPS

Figure 5-2. Waveform at TP1

5.Check the VCO Supply Voltage. Set the Calibrator to 1V at 1.2 MHz, operate. Using a DMM measure the dc voltage at U3 pin 8 and verify it is -5V ±7%. If a failure is detected, check Q2, Q3, U7, and associated components.

6.Check the VCO circuit. Power down the Calibrator, remove shorting header E6 from the J6 pins, and connect an external variable dc reference(2V to 13V) to TP2. Power up the Calibrator and set it for 1V at 1.2 Mhz, operate. Connect a frequency counter to TP3. Vary the external dc reference from 2 to 12V and verify the frequency counter reading changes from approximately 32 to 64 MHz. If a failure is detected, check the VCO circuit containing U3, CR1, CR2, and associated components.

7.Check the dividers in U4 and U5. Power down the Calibrator, remove shorting header E6 from the J6 pins, and connect an external variable dc reference (2 to 13V) to TP2. Power up the Calibrator and set it for 1V at 1.2 MHz, operate. Connect a

5-43

Page 335
Image 335
Fluke 5720A service manual 2. Waveform at TP1, F5-2.EPS