Fluke 5720A service manual After a few seconds, the following display appears

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 217
Image 217
After a few seconds, the following display appears:

Calibration and Verification 3

Calibrating the 5700A/5720A Series II to External Standards

8.Reverse the HI and LO connections at the 732B terminals, and press E. The following displays appear, asking you to wait before proceeding with the 10 kΩ standard.

PREV

MENU

IF3-8.EPS

After a few seconds, the following display appears:

PREV

MENU

IF3-9.EPS

9.Connect the calibrator to the 10 kΩ standard as shown in Figure 3-4 and enter the true value of the standard. If the standard is not between 9 kΩ and 11 kΩ, an error

message appears, which allows you to start over from this point with a different standard. Press Eagain to open the following display:

PREV

MENU

IF3-10.EPS

When the internal 10 kΩ reference has been characterized, the following message appears, which lets you accept or reject the changes that are about to be made to the calibration constant:

PREV

MENU

IF3-11.EPS

3-11

Page 217
Image 217
Fluke 5720A service manual After a few seconds, the following display appears