Table 3-2. List of Required Equipment for Main Output

5700A/5720A Series II Calibrator

Service Manual

Table 3-2. List of Required Equipment for Main Output

Equipment

AC Measurement Standard or Resistance Standards Reference Divider

Null Detector

Shunt Adapter

Calibrator

Frequency Counter

Shunts

High Current Shunt

Resistance Standard

DMM

Low Thermal Cables

Description

Fluke 5790A

Fluke 742A Series

Fluke 752A

EM Electronics N11 (Fluke 845A) Fluke 792A-7004

Fluke 55X0A

Philips PM6669

Fluke A40 Series

Fluke Y5020

L&N 0.1Ω 4221B

Wavetek 1281 or HP 3458A

Fluke 5440A-7002 (5700A) or 5440A-7003 (5720A or 5700A)

Application

ACV, ACI

Cal, Ohms, DCI

DC V

DC V

AC I

Ohms

Frequency

AC I

DC I

DC I

Ohms, DCI, ACI, AC V Various

Equipment Required for Wideband Ac Module (Option 5700-03) Verification

Equipment

Description

 

Application

AC Measurement Standard

Fluke 5790A with Wideband

Wideband

Option -03

 

Wideband Cable (supplied with 5700A-03)

50Ω Termination (supplied with

5700A-03)

Adapter (supplied with 5700A-03)

Fluke Cable

Wideband

Fluke Termination

Wideband

Pomona 1269 BNC(F) to dual

Wideband

banana plug

 

Adapter

Kings KN-99-46 N(F) to BNC(M)

Wideband

 

 

Equipment Required for Optional Tests

Equipment

Oscilloscope Mainframe

Differential Amplifier

Distortion Analyzer

Spectrum Analyzer

Description

 

Application

Tektronix 7000 Series

 

HF Noise

Tektronix 7A22

 

HF Noise

 

Krohn-Hite 6900B

 

AC V, ACI, Distortion

 

HP 8590A

 

AC V, Wideband

 

 

 

Distortion

 

 

 

Kelvin Varley Divider

Fluke 720A

DC V Linearity

 

 

3-24