3414 A12/A13: Nonlinear Control Loop 2vrng

Troubleshooting 5

Interpreting Diagnostic Fault Codes

within ± 100 mV. A failure indicates the Oscillator Control assembly (A12) is probably at fault.

3414 A12/A13: Nonlinear Control Loop 2vrng

The instrument is set for 1V 1 kHz operation. Signal ERROR INT. OUT on the Oscillator Control assembly is measured via the SDL line to be 0V ±5.0V. A failure indicates the Oscillator Control assembly (A12) is probably at fault.

3415 A12/A13: Nonlinear Control Loop 20vrng

The instrument is set for 10V 1 kHz operation. Signal ERROR INT. OUT on the Oscillator Control assembly is measured via the SDL line to be 0V ±5.0V. A failure indicates the Oscillator Control assembly (A12) is probably at fault.

3416 A12/A13: DAC 15th Bit Fault

The instrument is set for 1V 1 kHz operation. Bit 14 of the 15 BIT DAC is checked. A failure indicates the Oscillator Control assembly (A12) is probably at fault. Check FET Q19.

3417 A12: DC Sensor Buffer Fault

The instrument is set for 1V 1 kHz operation. The output of the DC Sense Buffer, on the Oscillator Control assembly, is measured via the RCL line. This measurement is done by comparing it to the DAC assembly output voltage. The difference is measured to be 0V ±500 μV. A failure indicates the Oscillator Control assembly (A12) is probably at fault.

3418 A12: Sensor Loop/SQ. Root Amp Fault

The instrument is set for 2.19V 1 kHz operation. The output of the DC Sense Buffer is connected to the main ac/dc sensor on the Oscillator Control assembly. The output of the main sensor is measured via the RCL line by comparing it to the DAC assembly output voltage. The difference is measured to be 0V ±100 mV. A failure indicates the Oscillator Control assembly (A12) is probably at fault.

3419 A12: AC Sensor Buff (2V Range)

The instrument is set for 2.19V 1 kHz operation. The output of the AC Sense Buffer is connected to the main ac/dc sensor on the Oscillator Control assembly. The output of the main sensor is measured via the RCL line by comparing it to the DAC assembly output voltage. The difference is measured to be 0V ±160 mV. A failure indicates the Oscillator Control assembly (A12) is probably at fault.

3420 A12: AC Sensor Buff (20V Range)

The instrument is set for 21.9V 1 kHz operation. The output of the AC Sense Buffer is connected to the main ac/dc sensor on the Oscillator Control assembly. The output of the main sensor is measured via the RCL line by comparing it to the DAC assembly output voltage. The difference is measured to be 0V ±160 mV. A failure indicates the Oscillator Control assembly (A12) is probably at fault.

3421 A12: AC Cal Sensor (2V Range)

The instrument is set for 2.19V 1 kHz and configured as in the 2V range ac/ac calibration. The output of the ac/ac sensor on the Oscillator Control assembly is

5-25