4. Enter the ambient temperature; then press E. The display shows:

Calibration and Verification 3

Calibrating the 5700A/5720A Series II to External Standards

4. Enter the ambient temperature; then press E. The display shows:

PREV

MENU

IF3-4.EPS

5. Connect the 732B to the calibrator as shown in Figure 3-3.

VOLTAGE REFERENCE

STANDARD

732B DC STANDARD

AC PWR

IN CAL

CHARGE

LOW BAT

10V

 

1.018V

CHASSIS

10V COM

1.018V COM

GUARD

SERIAL NUMBER

CALIBRATOR

OUTPUT SENSE

VΩA VΩ

HI

HI WIDEBAND

LO 5. Connect the 732B to the calibrator as shown in Figure 3-3.732B DC STANDARD LO

HI Figure 3-3. 732B External Calibration Connections.

AUX GUARD GROUND CURRENT

F7-3.EPS

Figure 3-3. 732B External Calibration Connections.

3-9

Page 215
Image 215
Fluke 5720A Enter the ambient temperature then press E. The display shows, Calibration and Verification, 732B DC STANDARD