Fluke 5720A service manual Press the “Cal “ softkey to bring up the following menu

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 224
Image 224
2. Press the “Cal “ softkey to bring up the following menu:

5700A/5720A Series II Calibrator

Service Manual

Since frequency flatness is determined by stable parameters (i.e. circuit geometry and dielectric constants), the flatness of the Wideband AC Module has excellent long-term stability. Consequently, a two-year calibration cycle is adequate for flatness calibration, and can be scheduled to coincide with the calibrator’s shipment to a standards laboratory for periodic verification.

The following procedure describes how to perform the wideband gain calibration.

Note

To perform this procedure you will need, in addition to the standard equipment supplied with the wideband option, a Type “N” female to double banana plug adapter (e.g., Pomona 1740).

Before you start this procedure, make sure the calibrator is powered on and has completed an appropriate warm-up period. Then proceed with this procedure to calibrate the wideband gain.

1.Press the “Setup Menus” softkey; then press the “Cal” softkey to being up the menu shown below:

PREV

MENU

IF3-1.EPS

2. Press the “Cal “ softkey to bring up the following menu:

PREV

MENU

IF3-2.EPS

3.Connect the wideband output cable between the WIDEBAND connector and the SENSE binding post.

The center conductor of the 50Ω feedthrough should go to SENSE HI and shown in Figure 3-6. The GND tab on the adapter should be on the LO side.

3-18

Page 224
Image 224
Fluke 5720A service manual Press the “Cal “ softkey to bring up the following menu