2-195. Current Guard Buffer

5700A/5720A Series II Calibrator

Service Manual

Internal calibration of the 2.2A range and gain of the High Voltage/High Current assembly is discussed further in the theory for the High Voltage assemblies.

2-195. Current Guard Buffer

Buffer amplifier U4, configured as a voltage follower, is used to provide a guard voltage equal to the output voltage across the external load. The guard voltage, if used, prevents any output current from being shunted away from the load due to leakage or shunt capacitance in the system cabling.

2-196. Compliance Limiter

A compliance limiter circuit consisting of Q24, Q25, and associated components clamps the output to ±11V during an over-compliance condition.

2-197. Current/Compliance Voltage Monitor

The current/compliance voltage monitor circuit, which contains CMOS analog switch U5A, U5C, U5D, op amp U6, and associated components, measures the voltage on either side of the current shunts. This allows the Calibrator to detect an over-current or over- compliance condition. A logic low on control line FET3 closes U5A to connect the monitor circuit to the input side of the shunt resistor. The measurement at this point is the sum of the output compliance voltage and voltage drop across the shunt, which is proportional to the output current. A logic high on control line FET3 opens U5C, which allows pull-down resistor R36 to close U5D. This connects the monitor circuit to the output side of the shunt resistor, which gives the output compliance voltage.

Op amp U6 and associated components create an absolute value circuit whose output, CUR/COMP MONITOR, is always a positive dc voltage. During operation in the ac current function U6B generates a negative half-wave signal equal to the positive peaks of its input. Resistors R31 and R30 sum this half-wave signal and the input signal at the input of U6A. Capacitor C14 averages the voltage so the output of U6A is a dc voltage which represents the average value of the selected input.

The diagnostic circuit connects CUR/COMP/ MONITOR to the SDL line, on which it is routed to the DAC assembly (A11) to be measured by the adc circuit. The calibrator software computes the difference between the two measurements and divides the result by the shunt value to determine the output current.

2-198. Current Assembly Calibration

Refer to Figure 2-36 for the following discussion. Internal calibration of the Current assembly is a process of determining the offset and gain constants for each of the four current ranges.

To determine the offset of the 220 mA range, the Current assembly is set to the positive dc 220 mA range with its input from the DAC assembly set to 0V. The output of the Current assembly is routed to the Ohms Main assembly (A10), via INT OUT HI, where it is connected to the previously calibrated 100Ω resistor.

To get a checkpoint reading, the Current assembly output is removed from the 100/Q resistor on Ohms (via the output switching relays). The 100Ω resistor is connected to a differential amplifier on the Ohms Cal assembly (A9). The output of this differential

2-126

Page 188
Image 188