13.The low-value (1Ω, 1.9Ω, and 10Ω) tests are summarized in Table 3-4.

5700A/5720A Series II Calibrator

Service Manual

DMM

INPUT

HI

SENSE

HI

LOLO

CURRENT SOURCE

742A-X

 

 

CALIBRATOR

OUTPUT

CURRENT

SENSE

OUTPUT

SENSE (UUT)

HI

HI

HI

HI

HI

LO

LO

LO

LO

LO

F3-5.EPS

Figure 3-8. 1 Ohm and 10 Ohm Resistor Verification

Note

There is no need to do the cal current calculation of step 6 if the current source has a settability of ±3 ppm.

12.Repeat steps 3 through 11 for the 1.9Ω and 10Ω resistance values using the 1.9Ω and 10Ω resistance standards. At the 10 ohm check, use 10 mA of current and a multiplier for step 3 of 0.01.

13.The low-value (1Ω, 1.9Ω, and 10Ω) tests are summarized in Table 3-4.