Table 3-24. 5700A AC Voltage Output Test Record (cont)

Calibration and Verification 3

Determining Test Limits for Other Calibration Intervals

Output

Level

Table 3-24. 5700A AC Voltage Output Test Record (cont)

 

Error

Spec

5790A

 

 

Frequency

1-Year

TUR

GB

Display

(± ppm )

Spec

 

Reading

 

Factor

 

 

(± ppm)

 

 

 

 

 

 

Test Limit (4)

(± ppm)

200V

1 kHz

 

85

31

2.74

0.971

83

200V

20 kHz

 

85

31

2.74

0.971

83

200V

50 kHz

 

240

69

3.48

0.991

238

200V

100 kHz

 

600

98

6.12

1.00

 

200V(2)

40 Hz

 

85

31

2.74

0.972

83

200V

20 Hz

 

185

68

2.72

0.970

179

200V

10 Hz

 

600

200

3.00

0.980

588

300V(3)

20 kHz

 

145

41

3.54

0.993

144

600V(3)

50 kHz

 

378

130

2.91

0.977

369

600V(3)

100 kHz

 

1375

500

2.21

0.945

1299

1 kV

1 kHz

 

84

38

2.21

0.945

79

1 kV

50 Hz

 

84

38

2.21

0.945

79

1 kV

300 Hz

 

84

38

2.21

0.945

79

1 kV(3)

20 kHz

 

131

38

3.45

0.991

130

1 kV(3)

30 kHz

 

371

130

2.85

0.975

362

1 kV(3)

40 Hz

 

84

38

2.21

0.945

79

1:This is a test of the bottom of the 20V range .

2:Observe the Calibrator output for 10 minutes and verify that it remains stable within ±7.5 ppm.

3:Perform only for units that are used with a 5725A Amplifier.

4:When the TUR (Test Uncertainty Ratio) is less than 4:1, the spec is reduced to give the same Consumer Risk as a 4:1 TUR as described later in this chapter.

Table 3-25. AC Voltage 2 mV Range Test Record

Characterized 1.9 mV Reading

Error Display Reading

 

 

Limits

±0.26%

3-61