16.Set the rear panel CALIBRATION switch to NORMAL.
IF3-13.EPS

Calibration and Verification 3

Calibrating the 5700A/5720A Series II to External Standards

11.Disconnect the 10 kΩ standard, and connect the calibrator to the 1Ω standard; then enter the true value of the 1Ω standard. If the standard is not between 0.9Ω and 1.1Ω, an error message appears, which lets you start over from this point with another standard. Press Eto bring up the following display:

PREV

MENU

When the internal 1Ω reference has been characterized, the following message appears, which lets you accept or reject the changes that re about to be made to the calibration constant:

PREV

MENU

IF3-14.EPS

12.To reject the changes, return to the calibration menu shown in step 2 by pressing P. Otherwise, press the softkey under “Proceed” to accept and save the changes, and let the calibrator complete the internal calibration steps.

13.The calibration is not effective until you store the newly calibrated constants in memory. To store the constants, set the rear panel CALIBRATION switch to ENABLE, the press the “Store Values” softkey.

Note

To review the proposed output shifts before you store the new constants, print a listing of the proposed shifts by pressing the softkey under “Print Output Shifts.”

14.After you store the constants, press the softkey under “DONE with Cal” to exit calibration and resume normal operation.

15.If you press this softkey before you store the constants, the new constants will be used temporarily for normal operation until the calibrator is powered down or reset. (This is only true for software versions G and lower. For versions H and higher, the process is aborted without updating existing constants.)

16.Set the rear panel CALIBRATION switch to NORMAL.

3-13

Page 219
Image 219
Fluke 5720A service manual Set the rear panel CALIBRATION switch to NORMAL