3822 Spare

5700A/5720A Series II Calibrator

Service Manual

3822 Spare

This fault code is not used.

3823 Spare

This fault code is not used.

3824 A8: Oven Regulation Fault

The heater voltage of the Switch Matrix HR1 hybrid assembly is measured to ensure it is warmed up and regulating by connecting OVEN TEMP to the SDL line where it is measured by the adc circuit on the DAC assembly. OVEN TEMP is measured to be 0.7V ±10% ±0.2V. A failure indicates the Switch Matrix (A8) assembly is probably at fault.

3825 Assembly A8 Not Responding

At power up the instrument checks to see what assemblies are not installed. This test verifies that the Switch Matrix assembly is not responding. A failure indicates the Switch Matrix (A8) assembly is probably at fault.

3826 A8: 8255 Was Reset

The control word of the 8255 on the Switch Matrix assembly is read. A failure indicates the Switch Matrix (A8) assembly is probably at fault.

3827 Assembly A8 Too Hot

3900 A5: 8255 Control Word

The diagnostics software reads the control word of the 8255 IC on the Wideband Output assembly. A failure indicates the Wideband Output (A5) assembly is probably at fault.

3901 A5: 8255 Port A Fault

The diagnostics software writes and reads data to port A of the 8255 IC on the Wideband Output assembly. A failure indicates the Wideband Output (A5) assembly is probably at fault.

3902 A5: 8255 Port B Fault

The diagnostics software reads port B of the 8255 IC on the Wideband Output assembly. A failure indicates the Wideband Output (A5) assembly is probably at fault.

3903 A5: 8255 Port C Fault

The diagnostics software reads port C of the 8255 IC on the Wideband Output assembly. A failure indicates the Wideband Output (A5) assembly is probably at fault.

3904 Optional Assemblies A5/A6 Are Missing

At power up the instrument checks to see what assemblies are not installed. This test verifies that the Wideband Output is not responding. A failure indicates the Wideband Output (A5) assembly is probably at fault.

5-36