Fluke 5720A Calibration and Verification, Determining Test Limits for Other Calibration Intervals

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 265
Image 265
Output

Calibration and Verification 3

Determining Test Limits for Other Calibration Intervals

Output

Level

Frequency

Table 3-23. 5720A AC Voltage Test Record (cont)

Error

Spec

5790A

 

 

1-Year

TUR

GB

Display

(± ppm )

Spec

Reading

 

Factor

 

(± ppm)

 

 

 

 

 

Test Limit (4)

(± ppm)

200V

1 kHz

 

64

31

2.05

0.936

59

200V

20 kHz

 

64

31

2.05

0.936

59

200V

50 kHz

 

101

69

1.46

0.875

88

200V

100 kHz

 

185

98

1.89

0.923

171

200V(2)

40 Hz

 

64

31

2.05

0.936

59

200V

20 Hz

 

110

68

1.62

0.896

99

200V

10 Hz

 

295

200

1.48

0.878

259

300V(3)

20 kHz

 

145

41

3.54

.993

144

600V(3)

50 kHz

 

378

130

2.91

.977

369

600V(3)

100 kHz

 

1375

500

2.21

.945

1299

1 kV

1 kHz

 

79

38

2.08

0.938

74

1 kV

50 Hz

 

79

38

2.08

0.938

74

1 kV

300 Hz

 

79

38

2.08

0.938

74

1 kV(3)

20 kHz

 

131

38

3.45

.991

130

1 kV(3)

30 kHz

 

371

130

2.85

.975

362

1 kV(3)

40 Hz

 

84

38

2.21

.945

79

1:This is a test of the bottom of the 20V range .

2:Observe the Calibrator output for 10 minutes and verify that it remains stable within ±7.5 ppm.

3:Perform only for units that are used with a 5725A Amplifier.

4:When the TUR (Test Uncertainty Ratio) is less than 4:1, the spec is reduced to give the same Consumer Risk as a 4:1 TUR as described later in this chapter.

3-59

Page 265
Image 265
Fluke 5720A Calibration and Verification, Determining Test Limits for Other Calibration Intervals, Output Level, Frequency