3-40. Determining Test Limits for Other Calibration Intervals

5700A/5720A Series II Calibrator

Service Manual

Table 3-15. Minimum Use Requirements (cont)

Item

No.

9.

10.

11.

12.

13.

14.

15.

16.

17.

18.

Description

Current Shunt Adapter

Frequency Counter

Standard Resistors (1)

DC Current Shunt (2)

AC/DC Current Shunt

Metal Film Resistors

Differential Amplifier

Distortion Analyzer

Kelvin-Varley Voltage Divider

HF Spectrum Analyzer (used in optional test for wideband distortion)

Minimum Use Specifications

Used in conjunction with 5790A and A40- series shunts to facilitate AC Current measurements

10 Hz to 30 MHz ±0.002%

0.1Ω nominal, true value certified to within 20 ppm, rated for 2A DC; 1Ω nominal, true value certified to within 6 ppm; 1.9Ω nominal, true value certified to within 6 ppm; 10Ω nominal, true value certified to within 6 ppm; 1 kΩ nominal, true value certified to within 5.5 ppm; 10 kΩ nominal, true value certified to within 3.5 ppm; 19 kΩ nominal, true value certified to within 4 ppm; 10 MΩ nominal, true value certified to within 15 ppm; 19 MΩ nominal, true value certified to within 28 ppm

Range: 10A

Uncertainty: ±0.008%

Ranges: 20 mA, 200 mA, 2A and 10A

Frequency: 10 Hz to 10 kHz

Uncertainty: ±310 ppm at 10 Hz; ±100 ppm at 20 Hz; ±50 ppm at 40 Hz, 1 kHz; ±100 ppm at 5 kHz, 10 kHz

Values: 200Ω, 2 kΩ, and 1 MΩ Temperature

C°: T9 or better Power Rating: 1/4 Watt

Tolerance: ±1%

Sensitivity: 5 μV rms Bandwidth selectable to 10 kHz

Range: 2V to 300V

Frequency: 10 Hz to 600 kHz

Ratio uncertainty: ±0.1 ppm of input

Freq. Range: 2 MHz to 120 MHz Input

Level: 3V (+20 dBm to -60 dBm)

Recommended

Equipment

Fluke 792A-7004

Philips PM 6669

Fluke 742A-1

Fluke 742A-1.9

Fluke 742A-10

Fluke 742A-1k

Fluke 742A-10k

Fluke 742A-19k

Fluke 742A-10M

Fluke 742A-19M

L&N 4221B (0.1Ω)

Fluke Y5020

Fluke A40-20 mA

Fluke A40-200 mA

Fluke A40-2A

Fluke A40A-10A (2)

Stock Items

Tektronix 7A22 w/7000- Series Mainframe

Krohn-Hite 6900B

Fluke 720A

HP 8590A

1:A DMM may be used for all but the 1Ω and 1.9Ω values. For those values using the DMM, a test method using an external current source is used for low-value resistance.

2:Needed only for 5725A Amplifier testing.

3-40. Determining Test Limits for Other Calibration Intervals

The verification procedures in this chapter test to the 90 day 99% confidence specification limits. For other calibration intervals it is necessary to calculate new specification limits and, if necessary, new test limits.

The following examples show how the 90 day limits were calculated. These examples illustrate how you can calculate the specifications and test limits, if necessary, for other calibration intervals.