4. Press the “Gain” softkey to bring up the following display:

Calibration and Verification 3

Calibrating the Wideband AC Module (Option 5700A-03)

CALIBRATOR

OUTPUT

SENSE

V Ω A

VΩ

50Ω FEEDTHROUGH

HI

WIDEBAND

HI

 

LO

LO

 

HI

 

 

AUX

GUARD GROUND

CURRENT

 

GND TAB

 

 

F7-5.EPS

Figure 3-6. Wideband Module Calibration Connection

4. Press the “Gain” softkey to bring up the following display:

PREV

MENU

IF3-3.EPS

5.Enter the ambient temperature; the press E. The display shows the following:

PREV

MENU

IF3-15.EPS

As the wideband calibration proceeds, messages appear on the display identifying all processes as they are encountered. When positive gains calibration is complete, a message appears telling you to refer to the manual for negative gains connections.