Fluke DC Current Specifications, Introduction and Specifications, 5720A 99%, 5720A 95%

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 51
Image 51
DC Current Specifications

Introduction and Specifications 1

DC Current Specifications

1-31. DC Current Specifications

Table 1-14. 5720A Series II DC Current Specifications: 99 % and 95 % Confidence Levels

5720A

99%

Range

99 % Confidence Level

 

 

Absolute Uncertainty

Resolution

±5 °C from calibration temperature.

 

For fields strengths >0.4 V/m but 3 V/m,

 

 

add 1 % of range.

Relative Uncertainty ±1 °C

220μA 2.2 mA

22mA

220mA

2.2 A [Note 1]

11 A

nA

0.1

1

10

μA

0.1

1

10

24 Hours

90 Days

180 Days

±(ppm output + nA)

40

+ 7

42

+ 7

45

+ 7

30

+ 8

35

+ 8

37

+ 8

30

+ 50

35

+ 50

37

+ 50

 

 

± (ppm output + μA)

40

+ 0.8

45

+ 0.8

47

+ 0.8

60

+ 15

70

+ 15

80

+ 15

 

 

 

 

 

 

5725A Amplifier:

 

 

 

 

 

 

 

 

330

+ 470

340

+ 480

350

+ 480

1 Year

50 + 7

40 + 8

40 + 50

50 + 0.8

90 + 15

360 + 480

24 Hours

90 Days

±(ppm output + nA)

24

+ 7

26

+ 7

24

+ 8

26

+ 8

24

+ 50

26

+ 50

±(ppm output + μA)

26

+ 0.5

30

+ 0.5

40

+ 12

45

+ 12

 

 

 

 

 

 

 

 

100

+ 130

110

+ 130

5720A

95%

Range

95 % Confidence Level

 

 

Absolute Uncertainty

Resolution

±5 °C from calibration temperature

 

For fields strengths >0.4 V/m but 3 V/m,

 

 

add 1 % of range.

 

 

 

Relative Uncertainty ±1 °C

220μA 2.2 mA

22mA

220mA

2.2 A [Note 1]

11 A

nA

0.1

1

10

μA

0.1

1

10

24 Hours

90 Days

180 Days

±(ppm output + nA)

32

+ 6

35

+ 6

37

+ 6

25

+ 7

30

+ 7

33

+ 7

25

+ 40

30

+ 40

33

+ 40

±(ppm output + μA)

35

+ 0.7

40

+ 0.7

42

+ 0.7

50

+ 12

60

+ 12

70

+ 12

 

 

 

 

 

 

5725A Amplifier:

 

 

 

 

 

330 + 470

340 + 480

350 + 480

1 Year

40 + 6

35 + 7

35 + 40

45 + 0.7

80 + 12

360 + 480

24 Hours

90 Days

±(ppm output + nA)

20

+ 6

22

+ 6

20

+ 7

22

+ 7

20

+ 40

22

+ 40

±(ppm output + μA)

20

+ 0.7

25

+ 0.7

32

+ 12

40

+ 12

 

 

 

 

100 + 130

110 + 130

Note: Maximum output from calibrator terminals is 2.2 A. Uncertainty specifications for 220 mA and 2.2 mA ranges are increased by a factor of 1.3 when supplied through 5725A terminals.

Specifications are otherwise identical for all output locations.

1.Add to uncertainty specifications:

±200 x I2 ppm for >100 mA on 220 mA range

±10 x I2 ppm for >1 A on 2.2 A range

1-27

Page 51
Image 51
Fluke DC Current Specifications, Introduction and Specifications, 5720A 99%, 5720A 95%, 99 % Confidence Level