Table 3-16. 5720A Resistance Test Record (cont)

Table 3-16. 5720A Resistance Test Record (cont)

Std

Res

Value

1.9MΩ (1)

10 MΩ)

19 MΩ

100MΩ

(2)

DMM Std Res Rdg

DMM Error

DMM

UUT Res

Value

UUT True Res Value

UUT Dev.

Uncert

Spec

Std

(ppm)

 

 

 

19

12.037

21.947

110

TUR

3.07

2.15

GB

Factor

0.982

0.942

Test

Limit

(±ppm)

36.3

44.3

Max

Difference of

Characterized to

Nominal Value

1.89962M to 1.90038M

9.997M to 10.003M

18.9943M to 19.0057M

99.95M to 100.05M

1:Not necessary to test due to 5720A Series II internal calibration process. Uncertainties and test limits are shown for other available 742A Standard Resistors.

2:Due to extremely slow settling time (approximately 5 minutes to 0.005% and sensitivity to any nearby movement, use of the DMM to test 100 megohms to the specified 0.01% uncertainty is not practical and therefore is not recommended. For those who wish to test it, a suitable way is to use an ESI SR 1050 10M/step Hamon-type Resistance Transfer Standard and use it in conjunction with an ESI 242-series bridge to effect the measurement to the required uncertainty.