Fluke 5720A Introduction and Specifications, DC Current Specifications, Overshoot 5 %

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 53
Image 53
Overshoot: <5 %

Introduction and Specifications 1

DC Current Specifications

Table 1-16. DC Current Secondary Performance Specifications and Operating Characteristics

Range

220μA 2.2 mA

22mA

220mA

2.2 A

5725A

11 A

Stability

±1 °C [Note 1]

24Hours

±(ppm output

+nA)

5 + 1

5 + 5

5 + 50

8 + 300

9 + 7 μA

±(ppm output

+μA)

25 + 100

Temperature

Coefficient [Note 2]

10 °-40 °C 0 °-10 °C and

40 °-50 °C

±(ppm output + nA)/ °C

01

+ 0.40

03

+ 1

01

+ 2

03

+ 10

01

+ 20

03

+ 100

01

+ 200

03

+ 1 μA

01

+ 2.5 μΑ

03

+ 10 μA

±(ppm output + μA)/ °C

20 + 75

30 + 120

 

 

Compliance

Limits

10

10

10

10

3 [Note 5]

4

Burden Voltage Adder [Note 3]

±nA/V

0.2

0.2

10

100

2μA

Maximum

Load for

Full

Accuracy

[Note 4]

Ω

20k

2k

200

20

2

4

 

Noise

 

 

 

Bandwidth

 

Bandwidth

0.1-10 Hz

10 Hz-10 kHz

 

 

 

pk-pk

 

RMS

ppm output

 

nA

 

+ nA

 

 

6

+ .9

10

6

+ 5

10

6

+ 50

50

9

+ 300

500

12

+ 1.5 μA

 

20 μA

ppm output

 

μA

 

+ μA

 

 

15

+ 70

175

 

 

 

 

Notes:

Maximum output from the calibrator’s terminals is 2.2 A. Uncertainty specifications for 220 mA and 2.2 mA ranges are increased by a factor of 1.3 when supplied through 5725A terminals.

1.Stability specifications are included in the Absolute Uncertainty values for the primary specifications.

2.Temperature coefficient is an adder to uncertainty specifications. It does not apply unless operating more than ±5 °C from calibration temperature.

3.Burden voltage adder is an adder to uncertainty specifications that does not apply unless burden voltage is greater than 0.5 V.

0.1x actual load

4.For higher loads, multiply uncertainty specification by: 1+

maximum load for full accuracy

5.The calibrator’s compliance limit is 2 V for outputs from 1 A to 2.2 A. 5725A Amplifier may be used in range- lock mode down to 0 A.

Minimum output: 0 for all ranges, including 5725A.

Settling time to full accuracy: 1 second for mA and mA ranges; 3 seconds for 2.2 A range; 6 seconds for 11 A range; + 1 second for range or polarity change

Overshoot: <5 %

1-29

Page 53
Image 53
Fluke 5720A service manual Introduction and Specifications, DC Current Specifications, Overshoot 5 %