5-5. Troubleshooting the Wideband Oscillator Assembly (A6)

5700A/5720A Series II Calibrator

Service Manual

voltages of 3V, 1V, 300 mV, and 30 mV, all at 1 kHz, and verify the DMM read the selected voltage ±5%. If a failure is detected, check K4-K8, Z1, and Z2. Next, connect the supplied output cable and 50Ω terminator to the WIDEBAND connector of the Calibrator front panel. Refer to the 5700A/5720A Series II Operators manual and run Wideband Gain calibration. After the Wideband Gain calibration is complete check the Wideband output at the end of the cable and 50Ω terminator at 3V, 1V, 300 mV, and 30 mV, all at 1 kHz, and verify the output is within 0.2%.

5-5. Troubleshooting the Wideband Oscillator Assembly (A6)

Proceed as follows to troubleshoot the Wideband Oscillator assembly (A5):

1.Remove the front and rear shields from the Wideband Oscillator assembly and place it up on the extender card. Power up the Calibrator and call up the Wideband function.

Note

All measurements are referenced to SCOM (TP16 or TP17) and all

Calibrator outputs are from the Wideband Option unless otherwise Noted.

2.Set the Calibrator to 1V at 1.2 MHz, operate. Connect an oscilloscope to TP10, set it to 500 mV/div at 500 ns/div, and verify it displays a1.2 MHz signal similar to that shown in Figure 5-1. Next, connect a frequency counter to TP10. Set the Calibrator to 2 MHz, 5 MHz, 10 MHz, 20 MHz, and 30 MHz, verify that each frequency is within 0.01% of the nominal. If a failure is detected, proceed with step 3 otherwise skip to step 11.

1 V

All measurements are referenced to SCOM (TP16 or TP17) and all 10 ms Manual background

F5-1.EPS

Figure 5-1. Waveform at TP10

3.Check the 8 MHz reference. Set the Calibrator to 1V at 1.2 MHz operate. Connect an oscilloscope to TP1 and verify it displays a 8 MHz signal similar to that shown in Figure 5-2. If a failure is detected, first verify that control line WB ON/OFF* is a

5-42