DC Voltage Specifications

Introduction and Specifications 1

DC Voltage Specifications

Table 1-3. 5700A Series II DC Voltage Specifications: 99 % and 95 % Confidence Levels

 

5700A

 

 

 

 

 

99%

99 % Confidence Level

 

 

 

 

 

 

 

 

 

 

 

 

 

Absolute Uncertainty

 

 

 

 

± 5 °C from calibration temperature

 

Range

 

Resolution

For fields strengths >1 V/m but 3 V/m,

 

 

 

 

 

add 0.01 % of range.

Relative Uncertainty

±1 °C

220mV

2.2 V

11V

22V

220V

1100 V

10 nV

100nV

1μV

1μV

10μV

100μV

24Hours

6.5+ .75

3.5+ 1.2

3.5+ 3

3.5+ 6

5+ 100

7+ 600

90 Days

180 Days

±(ppm output + μV)

7

+ .75

8

+ .75

6

+ 1.2

 

+ 1.2

7

5

+ 4

 

+ 4

7

5

+ 8

 

+ 8

7

6

+ 100

 

+ 100

8

8

+ 600

 

+ 600

10

1 Year

9 + .8

8 + 1.2

8 + 4

8 + 8

9 + 100

11 + 600

24 Hours

90 Days

±(ppm output + μV)

 

2.5

+ .5

4

+ .5

2.5

+ 1.2

 

+ 1.2

4

1.5

+ 3

 

+ 4

3.5

1.5

+ 6

 

+ 8

3.5

2.5

+ 100

 

+ 100

4

3

+ 600

 

+ 600

4.5

 

 

5700A

 

 

 

 

 

 

 

 

 

 

 

 

 

 

95%

95 % Confidence Level

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Absolute Uncertainty

 

 

 

 

 

 

 

 

 

 

± 5 °C from calibration temperature

 

 

 

Range

 

 

Resolution

 

For fields strengths >1 V/m but 3 V/m,

 

 

 

 

 

 

 

 

 

 

add 0.01 % of range.

 

 

 

 

 

 

 

 

 

 

24 Hours

 

90 Days

 

180 Days

 

1 Year

 

 

 

 

 

 

 

 

 

 

± (ppm output + μV)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Relative Uncertainty

±1 °C

24 Hours

90 Days

±(ppm output + μV)

220mV

2.2 V

11V

22V

220V

1100 V

10nV

100nV

1mV

1mV

10mV

100mV

5.5+ 0.6

3.5+ 1

3 + 3.5

3 + 6.5

4 + 80

6 + 500

6 + 0.6

5 + 1

4 + 3.5

4 + 6.5

5 + 80

7 + 500

7 + 0.6

6 + 1

6 + 3.5

6 + 6.5

7 + 80

8 + 500

8 + 0.6

7 + 1

7 + 3.5

7 + 6.5

8 + 80

9 + 500

2

+ 0.4

3.5

+ 0.4

2

+ 1

 

+ 1

3.5

1.2

+ 3

 

+ 3.5

3

1.2

+ 6

 

+ 7

3

2

+ 80

 

+ 80

3.5

2.4

+ 500

 

+ 500

4