5700A/5720A Series II Calibrator

Service Manual

switches to a high(~+17V) level while all the others remain at a low (-11V) level. If a failure is detected, check U8, U10, U11, and U13.

14.Check the Filters. Disconnect the cable between the Wideband Oscillator assembly and Wideband Output assembly at J1 on the Wideband Output assembly. Set the Calibrator to 1V at 1.2 MHz, 1.9 MHz, 2 MHz, 3.9 MHz, 4 MHz, 7.9 MHz, 8 MHz, 15.9 MHz, 16 MHz, and 30 Mhz. Connect an oscilloscope to TP14 and verify it displays a distortion-free sine wave at each frequency. If a failure is detected, check the corresponding Filter circuit.

5700A/5720A

Series II

Wideband Output

1.2Mhz

2.0Mhz

4.0Mhz

8.0Mhz

16.0Mhz

 

Table 5-1. Verifying Multiplexer U66

 

 

 

 

A

 

B

C

Pin 7

 

Pin 9

Pin 10

 

 

 

 

-5V

 

-5V

-5V

0V

 

-5V

-5V

 

-5V

 

0V

0V

 

-5V

 

-5V

0V

 

0V

 

0V

-5V

 

 

 

 

 

TP10

1-32 MHz

1.2MHz

2.0MHz

4.0MHz

8.0MHz

16.0MHz

1 V

F5-5.EPS 10 ms Manual background

F5-5.EPS

Figure 5-5. Waveform at Pin 6 of U9

5-46