Fluke 5720A service manual F5-9.EPS, F5-10.EPS

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 342
Image 342
F5-9.EPS

5700A/5720A Series II Calibrator

Service Manual

10 k Ω

 

HR2

1.2 Ω

SHUNT

 

R14

2V ac

 

±10%

K7C, K5D,

 

K10, K13

SCOM

F5-9.EPS

Figure 5-9. Verifying the 2.2 mA Range of the 220 μA /2.2 mA Amp Circuit

 

10 k Ω

 

 

 

HR2

1.2

Ω

 

SHUNT

 

R14

 

 

2V ac

 

 

 

±10%

 

K8C, K7A,

 

 

 

K5, K10, K13

 

SCOM

F5-10.EPS

Figure 5-10. Verifying the 22 mA Range of the 22 mA/220 mA Amp Circuit

7.Check the 220 mA Range of the 22 mA/220 mA AMP. Set the Calibrator to 200 mA, standby. Set the external ac reference to 4.6V (±0.1V) at 1 kHz. Using a DMM

measure the ac voltage at the collector of Q6 and verify it is 2.0V ac ±10%. If a failure first verify the 10Ω shunt on HR2 is correct and it's connected to SCOM through relays in K9C, K8C, K7A, K5D, K10, K13, and 1.2Ω load resistor R14 as shown in Figure 5-11. If the shunt and output switching are correct, check Q6-Q13, K4, and the associated components in the 22 mA/220 mA AMP circuit.

8.Check the NEG. FB BUFFER. Set the Calibrator to 200 mA, standby. Set the external ac reference to 4.6V at 1 kHz. Using a DMM measure the ac voltage at the collector of Q6 and Note the reading. Next, measure the voltage at TP3 and verify it is the same as the previous Noted reading ±0.01%. If a failure is detected, check U2, K5A, K6A, K8A,K9A and the associated components.

5-50

Page 342
Image 342