Fluke 5720A AC Current Test, 22 mA to 11A Ranges, Calibration and Verification, Full Verification

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 239
Image 239
5700A/5720A Series II Output

Calibration and Verification 3

Full Verification

5700A/5720A Series II Output

Level

Table 3-8. 5790A Adjustment Counts

5790A Display
5790A Adjustment Counts*

 

 

2 mV

2.0000

20 mV

20.0000

200 mV

200.0000

2V

2.000000

2.3V

2.30000

20V

20.00000

200V

200.0000

300V

300.000

600V

600.000

1000V

1000.000

*Adjustment counts of 3 times the listed value is allowed at 1 MHz

±1

±1

±4

±4

±1

±4

±4

±1

±2

±2

3-23. AC Current Test, 22 mA to 11A Ranges

This test requires the use of equipment listed in Table 3-9, when using the 5790A Input 1. Use Tables 3-33 (5720A) or 3-34 (5700A) for a test record.

Table 3-9. Equipment Required for 22 mA to 11A Alternating Current Test Using the 5790A Input 1

Equipment

AC Measurement standard Current Shunts

Current Shunt Adapter Cable

Cable (For 10A Setup)

Cable (For 10A Setup)

Adapter (For 10A Setup)

Adapter (For 10A Setup)

Adapter (For 10A Setup)

Model or Description

Fluke 5790A

Fluke A40 Series: 20 mA, 200 mA, 2A, and A40A Series: 10A (if verifying a 5725A) , with AC-DC difference corrections.

Fluke 792A-7004 A40 current shunt adapter

Pomona 1368-A-18. Double banana to single banana plugs. Pomona 5268-C-12. BNC(M) to single banana plugs.

Fluke A45-4003 (PN 212853) UHF(M) to UHF(M) with RG8A/U cable Pomona Model 1707. UHF(F) to banana adapter.

Kings KC-99-34 UHF(M) to BNC(F)

Pomona Model 1707. UHF(F) to banana adapter.

3-33