5-10. Troubleshooting the Ohms Cal Assembly (A9)

5700A/5720A Series II Calibrator

Service Manual

Note

This 2V can be as much as ±0.12V if the DAC assembly is uncalibrated. If a failure is detected, check the HR1 hybrid, U19, Q4, and associated components.

6.Check the mV Dividers. Set the Calibrator to 200 mV at 1 kHz, operate. Using a DMM measure the ac voltage at the cathode of CR6 (common to OUTPUT LO binding post) and verify it is 200 mV ±1%. Next, set the Calibrator to 20 mV at 1 kHz operate and again measure the ac voltage at the cathode of CR6. Verify it is 20 mV ±1%. If a failure is detected, check the resistive dividers in the HR1 resistor network and the associated relays.

5-10. Troubleshooting the Ohms Cal Assembly (A9)

Proceed as follows to troubleshoot the Ohms Cal assembly (A9):

1.The Ohms Cal assembly contains the two-wire lead drop compensation circuit, calibration circuitry, the 1Ω, 1.9Ω, and short resistance values. All relays on the Ohms Cal assembly are latching type. These relays are put in the set or reset position and they remain in that position even after power has been removed. To troubleshoot the path through the relays simply set the Calibrator to the function in which problem is detected and then turn off the Calibrator. The relays remain in either the set or reset position. Check individual relays by connecting an external power supply (a typical 9V battery works well) across the set coil to place the relay in the set position or the reset coil to place the relay in the reset position. Remove the rear shield from the assembly and put it up on the extender card. Power up the Calibrator and continue as follows.

2.Check the 1Ω, 1.9Ω, and short. Connect a four-terminal ohmmeter(such as a Fluke Model 8505A) to the Ohms Cal assembly as follows. Connect 8505A IN HI to TP1,

8505A IN LO to TP 2, 8505A SENSE HI to TP5, 8505A SENSE LO to TP6. Set the Calibrator to 1Ω, operate, EXT SNS, and verify the ohmmeter reads 1Ω ±0.1%. If a

failure is detected, check relays K4, K5, K6, and the four wire-wound resistors in R41. Set the Calibrator to 1.9Ω, operate, EXT SNS, and verify the ohmmeter reads 1.9Ω ±0.1%. If a failure is detected, check relays K4, K5, K30, and the two wire- wound resistors in R42. Set the Calibrator to 0Ω and verify the ohmmeter reads 0Ω ±0.1%. If a failure is detected, check relays K7 and K8.

5-11. Two-wire Compensation Circuit

Proceed as follows to troubleshoot the two-wire compensation circuit:

1.Check the +8A and -8A power supply for the two-wire comp. circuit. Set the Calibrator to 100Ω, operate, and turn the two-wire comp on. Using a DMM measure the dc voltage at TP8 (common to TP7) and verify it is +8.2V ±5%. Next, measure the dc voltage at TP9 (common to TP7) and verify it is -8.2V ±5%. If these supply voltages are correct skip to step 2. If a failure is detected, first connect an oscilloscope to pin 16 of U6 with the scope common connected to pin 13 of U6. Set the oscilloscope to 5V/div at 10 us/div and verify it displays a waveform similar the Figure 5-16. If a failure is detected, check U6, Q5, Q6, Control line PB7, and associated components. If this signal is correct then check T1, CR1, CR2, CR9, CR10, VR1, VR2, and associated components.

2.Set the Calibrator to 100Ω, operate, INT SNS, and 2-Wire Comp ON. Connect a Fluke 8505A DMM, set to the 100Ω range, across the Calibrator OUTPUT HI and OUTPUT LO binding posts. The current from the DMM should be 10 mA. Using another DMM measure the dc voltage between pins 4 and 5 of U7 and verify it is

5-56

Page 348
Image 348
Fluke 5720A service manual Troubleshooting the Ohms Cal Assembly A9, Two-wire Compensation Circuit