Table 1-19. 5700A Series II AC Current Specifications: 99 % Confidence Level

 

5700A

 

 

 

 

 

 

99%

99 % Confidence Level

 

 

 

 

 

 

 

 

 

 

 

 

 

Range

Resolution

Frequency

 

 

Absolute Uncertainty

 

±5 °C from calibration temperature

 

 

 

 

 

 

 

 

 

 

For fields strengths >0.4 V/m but 3 V/m,

 

 

 

 

 

 

add 1 % of range.

Relative Uncertainty

±1 °C

24 Hours

Hz

90 Days

180 Days

±(ppm output + nA)

1 Year

24 Hours

90 Days

±(ppm output + nA)

220μA

2.2 mA

22 mA

1 nA

10 nA

100 nA

10

- 20

650

+ 30

20

- 40

350

+ 25

40

- 1 k

120

+ 20

1k - 5 k

500

+ 50

05k - 10 k

1500

+ 100

10

- 20

650

+ 50

20

- 40

350

+ 40

40

- 1 k

120

+ 40

1k - 5 k

500

+ 500

05k - 10 k

1500

+ 1000

10

- 20

650

+ 500

20

- 40

350

+ 400

40

- 1 k

120

+ 400

1k - 5 k

500

+ 5000

05k - 10 k

1500

+ 10,000

 

 

 

 

700 + 30

380 + 25

140 + 20

600 + 50

1600 + 100

700 + 50

380 + 40

140 + 40

600 + 500

1600 + 1000

700 + 500

380 + 400

140 + 400

600 + 5000

1600 + 10,000

750 + 30

410 + 25

150 + 20

650 + 50

1700 + 100

750 + 50

410 + 40

150 + 40

650 + 500

1700 + 1000

750 + 500

410 + 400

150 + 400

650 + 5000

1700 + 10,000

800 + 30

420 + 25

160 + 20

700 + 50

1800 + 100

800 + 50

420 + 40

160 + 40

700 + 500

1800 + 1000

800 + 500

420 + 400

160 + 400

700 + 5000

1800 + 10,000

450 + 30

270 + 25

110 + 20

450 + 50

1400 + 100

450 + 50

270 + 40

110 + 40

450 + 500

1400 + 1000

450 + 500

270 + 400

110 + 400

450 + 5000

1400 + 10,000

500 + 30

300 + 25

120 + 20

500 + 50

1500 + 100

500 + 50

300 + 40

120 + 40

500 + 500

1500 + 1000

500 + 500

300 + 400

120 + 400

500 + 5000

1500 +

10,000

Hz

±(ppm output + μA)

±(ppm output + μA)

220 mA

2.2 A

1μA

10 μA

10

- 20

650

+ 5

20

- 40

350

+ 4

40

- 1 k

120

+ 4

1k - 5 k

500

+ 50

05k - 10 k

1500

+ 100

020 - 1 k

600

+ 40

01 k - 5 k

700

+ 100

05 k - 10 k

8000

+ 200

700 + 5

380 + 4

150 + 4

600 + 50

1600 + 100

650 + 40

750 + 100

9000 + 200

750 + 5

410 + 4

170 + 4

650 + 50

1700 + 100

700 + 40

800 + 100

9500 + 200

800 + 5

420 + 4

180 + 4

700 + 50

1800 + 100

750 + 40

850 + 100

10,000 + 200

450 + 5

280 + 4

110 + 4

450 + 50

1400 + 100

600 + 40

650 + 100

7500 + 200

500 + 5

300 + 4

130 + 4

500 + 50

1500 + 100

650 + 40

750 + 100

8500 + 200

11 A

100μA

5725A Amplifier:

040 - 1 k

370

+ 170

400 + 170

440

+ 170

460

+ 170

01 k - 5 k

800

+ 380

850 + 380

900

+ 380

950

+ 380

05 k - 10 k

3000

+ 750

3300 + 750

3500

+ 750

3600

+ 750

300 + 170

700 + 380

2800 + 750

330 + 170

800 + 380

3200 + 750

Note: Maximum output from the calibrator’s terminals is 2.2 A. Uncertainty specifications for 220 μA and 2.2 mA ranges are increased by 1.3 x plus 2 μA when supplied through 5725A terminals. Specifications are otherwise identical for all output locations.