Fluke 5720A Calibration and Verification, Determining Test Limits for Other Calibration Intervals

Models: 5720A

1 570
Download 570 pages 1.07 Kb
Page 263
Image 263
Table 3-21. Direct Current Accuracy Test Record (5700A)

Calibration and Verification 3

Determining Test Limits for Other Calibration Intervals

Table 3-21. Direct Current Accuracy Test Record (5700A)

Output Current

+2A -2A

+200 mA -200 mA +20 mA -20 mA +2 mA -2 mA +200 μA -200 μA

Std

Res

Value

0.1Ω

0.1Ω

1.0Ω

1.0Ω

10Ω

10Ω

1kΩ

1kΩ

10 kΩ

10 kΩ

Characterizing dc DMM @ ±200 mV and ±2.0V

DC DMM

Rdg

Column A

Column B

UUT Actual

Std Res

UUT Error

Error (A+B)

Setting

 

Correction

Reading

 

(ppm)

 

 

(ppm)

(ppm)

 

 

 

(200 mV)

(200 mV)

(200 mV)

(200 mV)

(200 mV)

(200 mV)

(2V)

(2V)

(2V)

(2V)

5725A Amplifier DC Current Test

Test Limit,

90-day

(ppm)

135

135

73

73

55

55

55

55

100

100

Output Current

+10A

-10A

+5A

+3A

-3A

DC DMM Reading

Certified Shunt

Value

Cal Actual

Current

Test Limits, 90-

day

(±ppm)

388ppm

388ppm

436ppm

500ppm

500ppm

Table 3-22. AC Voltage Frequency Accuracy Test Record

Frequency

10Hz

15Hz

100Hz

200Hz

500Hz

1kHz

5kHz

10kHz

140kHz

200kHz

500kHz

1MHz

Tolerance

99.99ms - 100.01 ms 66.673 ms - 66.66 ms 9.999 ms - 10.001 ms

199.98Hz - 200.02 Hz

499.95Hz - 500.05 Hz

999.9Hz - 1000.1 Hz 4999.5 Hz - 5000.5 Hz 9.999 kHz - 10.001 kHz 139.986 kHz - 140.014 kHz

199.98kHz - 200.02 kHz

499.95kHz - 500.05 kHz

0.9999 MHz - 1.0001 MHz

Actual

3-57

Page 263
Image 263
Fluke 5720A Calibration and Verification, Determining Test Limits for Other Calibration Intervals, Output Current, Dc Dmm