SITRANS P measuring instruments for pressure

 

 

Remote seals

 

 

Measuring setups without remote seals

 

 

 

 

Also you must make sure that the level in the container is al-

Overview

ways above the top spigot.

 

 

Notes

• When measuring density, make sure that the level of the me-

• For the separation layer measurement, the separation layer

dium remains constant. The level should be above the top

spigot

 

has to be positioned between the two spigots.

2

 

 

 

 

 

 

 

Dimensional drawings

 

 

 

 

 

 

 

Pressure transmitters for differential pressure, for flanging

 

 

 

 

Measuring setups for open containers

 

 

 

 

 

 

 

 

Full-scale value

H o

 

r

 

 

 

Start-of-scale value

 

 

 

 

 

 

H u

 

 

 

 

+

 

 

 

 

 

 

 

 

 

Constant level

 

 

 

 

 

 

 

 

 

= full-scale value

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

r2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Separation layer

H

 

o

 

 

r1

 

 

 

 

 

 

 

 

 

 

Start-of-scale value

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Hu

+

Level measurement

Start-of-scale:

pMA = r · g · HU

Full-scale:

pME = r · g · HO

Legend

 

pMA

Start-of-scale value to be set

pME

Full-scale value to be set

rDensity of medium in vessel

gLocal acceleration due to gravity

HU

Start-of-scale value

HO

Full-scale value

Separation layer measurement

Start-of-scale: pMA = g · (HU · r1 + (HO - HU) · r2)

Full-scale:

pME = r1 · g · HO

Legend

 

pMA

Start-of-scale value to be set

pME

Full-scale value to be set

r1

Density of heavier liquid

r2

Density of lighter liquid

gLocal acceleration due to gravity

HU

Start-of-scale value

HO

Full-scale value

Ho

Constant level

r

Density measurement

Start-of-scale: pMA = rMIN · g · HO

Full-scale:

pME = rMAX · g · HO

Legende

 

pMA

Start-of-scale value to be set

pME

Full-scale value to be set

rMIN

Minimum density of medium in vessel

rMAX

Maximum density of medium in vessel

gLocal acceleration due to gravity

HO

Full-scale value in m

+

Siemens FI 01 · 2009