Siemens P technical specifications Transmitters for food, pharmaceuticals and biotechnology

Models: P

1 234
Download 234 pages 4.66 Kb
Page 22
Image 22

© Siemens AG 2008

SITRANS P measuring instruments for pressure

Transmitters for food, pharmaceuticals and biotechnology

SITRANS P Compact

for gage and absolute pressure

 

 

Function

 

 

 

 

 

The process pressure acts on a piezo-resistive semiconductor

 

 

 

 

measuring bridge through a remote seal and a transmission liq-

 

 

 

 

uid. The pressure transmitter converts the pressure values into a

 

2

 

 

 

load-independent current.

 

 

 

 

A compensation network makes the output signal largely inde-

 

 

 

 

pendent of the ambient temperature. As a result of a specially

 

 

 

 

adapted remote seal connection with minimized volume, the in-

 

 

 

 

fluence of the process temperature on the output signal is

 

 

 

 

greatly reduced compared to a conventional screw connection.

 

 

 

 

 

The pressure transmitters can be powered with a non-regulated

 

 

 

 

DC voltage of 10 to 30 V. Output signals common to measuring

 

 

 

 

technology are available.

 

 

 

 

 

 

 

 

 

Technical specifications

 

 

 

 

 

 

 

 

 

 

Pressure transmitters for food, pharmaceuticals and biotechnology

 

 

 

 

 

 

 

 

 

 

Mode of operation

 

 

 

 

 

Measuring principle

Piezo-resistive

 

 

 

 

 

 

 

 

 

 

Input

 

 

 

 

 

Measured variable

Gage or absolute pressure

 

 

 

 

Measured range

0 ... 160 mbar (0 ... 2.32 psi)

 

 

 

 

 

...

 

 

 

 

 

0 ... 40 bar (0 ... 580 psi)

 

 

 

 

 

 

 

 

 

 

Output

 

 

 

 

 

Output signal

 

 

 

 

 

Two-wire system

4 ... 20 mA

 

 

 

 

Three-wire system

0 ... 20 mA

 

 

 

 

 

 

 

 

 

 

Measuring accuracy

To EN 60770-1

 

 

 

 

Linearity error including hysteresis

0.2% of full-scale value

 

 

 

 

(reference point adjustment)

 

 

 

 

 

Adjustment accuracy

≤ ± 0.2% of full-scale value

 

 

 

 

Adjustment time

< 20 ms

 

 

 

 

Influence of ambient temperature

 

 

 

 

 

On th enclosure

 

 

 

 

 

• Zero

< 0.2%/10 K of full-scale value

 

 

 

 

• Measured span

< 0.2%/10 K of full-scale value

 

 

 

 

On the process connection (remote

Zero error (depends on design)

 

 

 

 

seal)

 

 

 

 

 

• Flange remote seal

 

 

 

 

 

- DN 25 / 1"

4.8 mbar/10 K (0.070 psi/10 K)

 

 

 

 

- DN 32 / 1¼”

2.3 mbar/10 K (0.033 psi/10 K)

 

 

 

 

- DN 40 / 1½”

1.6 mbar/10 K (0.023 psi/10 K)

 

 

 

 

- DN 50 / 2"

0.6 mbar/10 K (0.009 psi/10 K)

 

 

 

 

Clamp-on seal

 

 

 

 

 

- DN 25 / 1"

9.5 mbar/10 K (0.138 psi/10 K)

 

 

 

 

- DN 32 / 1¼”

4.1 mbar/10 K (0.060 psi/10 K)

 

 

 

 

- DN 40 / 1½”

3.9 mbar/10 K (0.057 psi/10 K)

 

 

 

 

- DN 50 / 2"

3.9 mbar/10 K (0.057 psi/10 K)

The zero error specified for the process connection should be consid- ered as a guideline for a standard design. We will produce a detailed system calculation on request. Systems with reduced remote seal errors are available on request.

Rated conditions

 

Installation conditions

 

• Mounting position

Any, vertical as standard

Ambient conditions

 

• Ambient temperature

-10 ... +70 °C (14 ... 158 °F)

• Storage temperature

-10 ... +90 °C (14 ... 194 °F)

• Process temperature

Max. 200 °C (392 °F), depends

 

on design

• Degree of protection (to EN 60529)

IP65, optional IP67

• Electromagnetic compatibility

 

- Emitted interference

To EN 50081 Part 1, issue 1993

 

(residential and industrial areas).

 

The unit has no own emissions.

- Interference immunity to

EN 50082 Part 2, issue March

 

1995 (industrial areas)

 

 

Design

 

Weight (without remote seal)

 

• Field housing

460 g (1.01 lb)

• Housing with plug

200 g (0.44 lb)

Housing

 

• Designs

• Field housing IP65 or IP67, with

 

screwed gland

 

• Angled plug DIN 43650, IP65

 

• Cable connection, IP67

 

• Round plug connector M12,

 

IP65

• Material

Stainless steel, mat.

 

No. 1.4404/1.4305

Material of union nut

Polyamide (with electrical con-

 

nection using plug or cable)

 

Electronics unit potted with sili-

 

cone

 

Internal ventilation for measuring

 

ranges < 16 bar (< 232 psi),

 

through housing thread or con-

 

nection cable depending on

 

design

Process connection

 

• Versions

See Ordering data

• Material of coupling

Stainless steel, mat.

 

No. 1.4404/316L

 

 

Power supply

 

Terminal voltage on transmitter

10 ... 30 V DC

Rated voltage

24 V DC

 

 

Certificate and approvals

 

Classification according to pressure

For gases of fluid group 1 and liq-

equipment directive

uids of fluid group 1; complies

(DRGL 97/23/EC)

with the requirements of article 3,

 

paragraph 1 (appendix 1);

 

assigned to category III, confor-

 

mity evaluation module H by the

 

TÜV Nord

Explosion protection

 

• Intrinsic safety "i"

TÜV 03 ATEX 2099 X

- Identification

Ex II 2G EEx ib IIC T6

2/22

Siemens FI 01 · 2009

Page 22
Image 22
Siemens P technical specifications Transmitters for food, pharmaceuticals and biotechnology