Siemens +150 C, Design standard version, Design version with front-flush dia Phragm

Models: P

1 234
Download 234 pages 4.66 Kb
Page 36
Image 36

© Siemens AG 2008

SITRANS P measuring instruments for pressure

Transmitters for food, pharmaceuticals and biotechnology

SITRANS P300 for gage and absolute pressure

 

 

SITRANS P300 for gage pressure and absolute pressure

 

 

 

 

 

 

 

 

 

HART

 

PROFIBUS PA and FOUNDATION Fieldbus

 

 

 

 

 

 

 

 

 

Rated operating conditions

 

 

 

 

 

 

Installation conditions

 

 

 

 

2

 

 

 

 

 

Ambient temperature

Observe the temperature class in areas subject to explosion hazard.

 

 

 

 

• Measuring cell with silicone oil

-40 ... +85

°C (-40 ... +185 °F)

 

 

• Measuring cell with Neobee oil (with front-

-10

... +85

°C (14 ... +185 °F)

 

 

flush diaphragm)

 

 

 

 

 

 

• Measuring cell with inert liquid (not with

-20

... +85

°C (-4 ... +185 °F)

 

 

 

front-flush diaphragm)

 

 

 

 

 

 

• Digital display

-30

... +85

°C (-22 ... +185 °F)

 

 

• Storage temperature

-50

... +85

°C (-58 ... +185 °F)

 

 

 

(with Neobee: -20 ... +85 °C (-4 ... +185 °F)

 

 

Climatic class

 

 

 

 

 

 

Condensation

Permissible

 

 

Degree of protection to EN 60529

IP65, IP68, NEMA X, enclosure cleaning, resistant to lyes, steam to 150° C (302 °F)

 

 

Electromagnetic compatibility

 

 

 

 

Emitted interference and interference immunity To EN 61326 and NAMUR NE 21 Medium conditions

Process temperature

• Measuring cell with silicone oil

-40 ...

+100 °C (-40 ...

+212 °F)

• Measuring cell with silicone oil (with front-

-40 ...

+150 °C (-40 ...

+302 °F)

flush diaphragm)

 

 

 

• Measuring cell with Neobee oil (with front-

-40 ...

+150 °C (-40 ...

+302 °F)

flush diaphragm)

 

 

 

Measuring cell with silicone oil, with temper- -40 ... +200 °C (-40 ... +392 °F) ature isolator (only with front-flush dia-

phragm)

• Measuring cell with inert liquid

-20 ... +100 °C (-4 ... +212 °F)

• Measuring cell with high temperature oil

-10 ... +250 °C (14 ... +482 °F)

 

 

Design (standard version)

 

Weight (without options)

Approx. 800 g (1.8 lb)

Housing material

Stainless steel, mat. No. 1.4301/304

Material of parts in contact with the medium

 

• Connection shank

Stainless steel, mat. No. 1.4404/316L or Hastelloy C276, mat. No. 2.4819

• Oval flange

Stainless steel, mat. No. 1.4404/316L

• Seal diaphragm

Stainless steel, mat. No. 1.4404/316L or Hastelloy C276, mat. No. 2.4819

• Measuring cell filling

• Silicone oil

 

• Inert filling liquid

Process connection

• G½B to DIN EN 837-1

 

• Female thread ½-14 NPT

 

• Oval flange PN 160 (MWP 2320 psi) with fastening thread:

 

- 7/ -20 UNF to IEC 61518

 

16

 

- M10 as per DIN 19213

 

 

Design (version with front-flush dia-

 

phragm)

 

Weight (without options)

Approx. 1 ... 13 kg (2.2 ... 29 lb)

Housing material

Stainless steel, mat. No. 1.4301/304

Material of parts in contact with the medium

 

• Process connection

Stainless steel, mat. No. 1.4404/316L

• Seal diaphragm

Stainless steel, mat. No. 1.4404/316L

• Measuring cell filling

• Silicone oil

 

• Inert filling liquid

 

• FDA compliant fill fluid (Neobee oil)

Process connection

• Flanges as per EN and ASME

 

• F&B and pharmaceutical flanges

2/36

Siemens FI 01 · 2009

Page 36
Image 36
Siemens technical specifications +150 C, Design standard version, Design version with front-flush dia Phragm