SITRANS P measuring instruments for pressure

 

 

 

 

Transmitters for food, pharmaceuticals and biotechnology

 

 

 

 

 

 

 

SITRANS P300 for gage and absolute pressure

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SITRANS P300 for gage pressure and absolute pressure

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HART

 

 

 

 

 

 

PROFIBUS PA and FOUNDATION Fieldbus

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Input of absolute pressure, with front-flush diaphragm

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measured variable

 

Absolute pressure (front-flush)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

Spans (infinitely adjustable) or

 

Span

 

Max. perm. test pres-

 

Nominal measuring

 

Max. perm. test pres-

 

 

 

 

 

 

nominal measuring range and

 

 

 

sure

 

 

range

 

sure

 

 

 

 

max. pemissible test pressure

 

43 ... 1300 mbar a

 

10 bar a

 

 

1300 mbar a

 

10 bar a

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(0.62 ... 18.9 psi a)

 

(145 psi a)

 

 

(18.9 psi a)

 

(145 psi a)

 

 

 

 

 

 

 

0.16 ... 5 bar a

 

30 bar a

 

 

5 bar a

 

30 bar a

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(2.32 ... 72.5 psi a)

 

(435 psi a)

 

 

(72,5 psi a)

 

(435 psi a)

 

 

 

 

 

 

 

1 ... 30 bar a

 

100 bar a

 

 

30 bar a

 

100 bar a

 

 

 

 

 

 

 

(14.5 ... 435 psi a)

 

(1450 psi a)

 

 

(435 psi a)

 

(1450 psi a)

 

 

 

 

 

 

 

Depending on the process connection, the span

 

Depending on the process connection, the nomi-

 

 

 

 

may differ from these values

 

 

nal measuring range may differ from these values

 

 

Lower measuring limit

 

0 bar a (0 psi a)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Upper measuring limit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

• Measuring cell with silicone oil

 

100% der max. Messspanne

 

 

100% des max. Nennmessbereichs

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Output

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Output signal

 

4 ... 20 mA

 

 

 

 

 

 

Digital PROFIBUS PA signal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Physical bus

 

-

 

 

 

 

 

 

IEC 61158-2

 

 

 

 

 

 

 

With polarity reversal protection

 

No

 

 

 

 

 

 

Yes

 

 

 

 

 

 

 

Electrical damping T63 (step width 0.1 s)

 

Set to 0.1 s (0 ... 100 s)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Accuracy

 

To EN 60770-1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Reference conditions

 

Increasing characteristic, start-of-scale value 0 bar, stainless steel seal diaphragm, measuring cell

 

 

(All error data refer always refer to the set

 

with silicone oil, room temperature 25 °C (77 °F), span ratio (r = max. span / set span)

 

 

 

 

 

span)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Measurement deviation with cut-off point set-

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ting, including hysteresis and repeatability.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Gage pressure

 

Absolute

 

(Absolute

 

Gage pressure

 

Absolute

 

(Absolute

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

pressure

 

pressure,

 

 

 

pressure

 

pressure,

 

 

 

 

 

 

 

 

 

front-

 

 

 

 

 

front-

 

 

 

 

 

 

 

 

 

flushed)

 

 

 

 

 

flushed)

 

 

Linear characteristic curve

 

 

 

 

 

 

 

 

0,075%

 

0,1%

 

0,2%

 

 

• r 10

 

(0.0029 r + 0.071)%

 

0,1%

 

0,2%

 

 

 

 

 

 

 

 

 

• 10 < r 30

 

(0.0045 r + 0.071)%

 

0,2%

 

0,4%

 

 

 

 

 

 

 

 

 

• 30 < r 100

 

(0.005 r + 0.05)%

 

-

 

 

-

 

 

 

 

 

 

 

 

 

Settling time T63 without electrical damping

 

Approx. 0.2 s

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Long-term drift at ± 30 °C (± 54 °F)

 

(0.25 r)%/5 years

 

(0.1 r)%/year

 

0.25%/5 years

 

0.1%/year

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Influence of ambient temperature

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

• at -10 ... +60 °C (14 ... 140 °F)

 

(0.1 r +0.2)%

 

 

 

 

(0.2 r

 

0,3%

 

 

 

0,5%

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+0.3)%

 

 

 

 

 

 

 

 

 

• at -40 ... -10 °C and +60 ... +85 °C

 

(0.1 r + 0.15)% / 10 K

 

 

 

 

(0.2 r +

 

0.25%/10 K

 

 

 

0.5%/

 

 

(-40 ... 14 °F and 140 ... 185 °F)

 

 

 

 

 

 

0.3)%/10 K

 

 

 

 

 

10 K

 

 

Influence of the medium temperature (only

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

with front-flush diaphragm)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

• Temperature difference between medium

 

3 mbar/10 K (0.04 psi/10 K)

 

 

 

 

 

 

 

 

 

 

temperature and ambient temperature

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Siemens FI 01 · 2009

2/35

Page 35
Image 35
Siemens P ≤ 0,1% ≤ 0,2% 10 r ≤, ≤ 0,2% ≤ 0,4% 30 r ≤ ≤ 0.005 ⋅ r + 0.05%, ≤ 0.2 ⋅ r + ≤ 0.25%/10 K ≤ 0.5% 140 .. F