© Siemens AG 2008

SITRANS P measuring instruments for pressure

Transmitters for gage, absolute and differential pressure

ZD series for gage and absolute pressure

Upper and lower current saturation limit

Electrical damping

The other two keys are used to set the values in the individual op- erating modes.

2Two LED indicators are fitted above the display to monitor the set range and the status.

The green LED signals that the measured pressure lies within the set limits. The red LED lights up when the measured pressure lies outside the set limits and when there is an error.

Technical specifications

SITRANS P pressure transmitters, ZD series

Mode of operation

 

Measuring principle

Thin-film strain gage

 

 

Input

 

Measured variable

Gage and absolute pressure

Measured range

Resolution

0

... 2 bar (0 ... 29 psi)

0.6 mbar (0.008 psi)

0

... 10 bar (0 ... 145 psi)

3 mbar (0.044 psi)

0

... 50 bar (0 ... 725 psi)

15 mbar (0.218 psi)

0

... 200 bar (0 ... 2900 psi)

60 mbar (0.9 psi)

0

... 400 bar (0 ... 5800 psi)

120 mbar (1.8 psi)

Measured range

Overload limit

0

... 2 bar (0 ... 29 psi)

5 bar (72.5 psi)

0

... 10 bar (0 ... 145 psi)

25 bar (363 psi)

0

... 50 bar (0 ... 725 psi)

120 bar (1740 psi)

0

... 200 bar (0 ... 2900 psi)

500 bar (7250 psi)

0

... 400 bar (0 ... 5800 psi)

600 bar (8700 psi)

Range adjustment (turndown)

5:1

 

 

Output

 

Output signal

4 ... 20 mA

Lower current limit

min. 3.6 mA

Upper current limit

max. 23 mA

Output protected against

Reversed polarity, overvoltage

 

 

and short-circuiting

Max. load

R B = (UH - 12 V) / 0.023 A

Voltage measurement

Linear rising

 

 

Measuring accuracy

To EN 60770-1

Error in measurement (including

< 0.25% of full-scale value (typi-

non-linearity, hysteresis and repeat-

cal), max. 0.5%

ability, at 25 °C (77 °F))

 

Adjustment time

< 100 ms

Long-term drift

0.25% of full scale value/year

Influence of ambient temperature

< ±0.25%/10 K (< ±0.25%/10 K)

 

 

of full-scale value

Vibration influence

0.05%/g to 500 Hz in all directions

 

 

(to IEC 68-2-64)

Power supply effect

< ±0.01%/V of full-scale value

Rated conditions

 

Ambient conditions

 

• Ambient temperature

-25 ... +85 °C (-13 ... +185 °F)

• Storage temperature

-40 ... +85 °C (-40 ... +185 °F)

Medium conditions

 

• Process temperature

-30 ... +100 °C (-22 ... +212 °F)

Degree of protection

IP65 to EN 60529

Electromagnetic compatibility

 

• Emitted interference and interfer-

To EN 61326/A1 appendix A

ence immunity

(1998)

 

 

Displays and controls

 

Display

LCD, max. 5 digits, digit height

 

9 mm

Decimal point

Freely parameterizable

Limit values

Freely parameterizable

Limit violation display

Red LED and message on LCD

 

(symbol /symbol in case of

 

limit violation in upward / down-

 

ward direction)

Parameterization

With 3 keys

Units

mA or % or physical variable

 

(default setting: bar)

 

Other dimensions: mbar, kPa,

 

MPa, mmH20, mH20, psi, inH20,

 

mmHg, kg/cm², torr, atm

Damping

Between 0.1 and 100 s

(increment: 0.1 s) freely parame-

 

 

terizable

 

 

Design

 

Weight

0.6 kg (1.32 lb)

Electrical connection

Using 2-pole plug connector with

 

M16x1.5-Cable inlet to

 

EN 175301-803A, plastic

Process connection

• Male thread G½B and female

 

thread G1/ B

 

8

 

• G½B to EN 837-1

 

• Female thread: ½-14 NPT

Version of housing/process connec-

• Radial (type A), can be swiveled

tion

by max. ±120° (α)

 

• Axial (type B), can be swiveled

 

by max. ±360°

Material

 

Non-wetted parts materials

 

• Field housing

Ø 80 mm (3.15 inch), stainless

 

steel mat. No. 1.4016

• Cover

Stainless steel, mat. No. 1.4016

 

with glass

Wetted parts materials

 

• Measuring cell

Al2O3

• Gasket

Viton

• Process connection

Stainless steel, mat. No.

 

1.4571/316Ti

 

 

Power supply

 

Terminal voltage on pressure trans-

12 ... 30 V DC

mitter (UH)

 

Certificate and approvals

 

Classification according to pressure

For gases of fluid group 1 and liq-

equipment directive 97/23/EC

uids of fluid 1; complies with

 

requirements of article 3, para-

 

graph 3 (sound engineering prac-

 

tice)

2/18

Siemens FI 01 · 2009

Page 18
Image 18
Siemens technical specifications Displays and controls, Power supply