Siemens technical specifications 124, Sitrans P, DS III series for level

Models: P

1 234
Download 234 pages 4.66 Kb
Page 124
Image 124

© Siemens AG 2008

SITRANS P measuring instruments for pressure

Transmitters for gage, absolute and differential pressure, flow and level

DS III series for level

Technical specifications

SITRANS P, DS III series for level

 

 

 

 

HART

 

 

 

PROFIBUS PA or FOUNDATION Fieldbus

 

 

 

 

 

 

 

 

 

 

 

2

 

Input

 

 

 

 

 

 

 

 

 

Measured variable

 

Level

 

 

 

 

 

 

 

 

Spans (infinitely adjustable) or

Span

 

Maximum working

 

Nominal measuring

 

Maximum working

 

 

 

 

 

 

 

nominal measuring range and

 

 

pressure

 

range

 

pressure

 

 

max. permissible working pressure

25 ... 250 mbar g

 

See "Mounting flange"

 

250 mbar g

 

See "Mounting flange"

 

 

 

 

 

 

 

 

 

 

 

(0.36 ... 3.63 psi g)

 

 

 

(3.63 psi g)

 

 

 

 

 

 

25 ... 600 mbar g

 

See "Mounting flange"

 

600 mbar g

 

See "Mounting flange"

 

 

 

 

(0.36 ... 8.7 psi g)

 

 

 

(8.7 psi g)

 

 

 

 

 

 

53 ... 1600 mbar g

 

See "Mounting flange"

 

1600 mbar g

 

See "Mounting flange"

 

 

 

 

(0.77 ... 23.2 psi g)

 

 

 

(23.2 psi g)

 

 

 

 

 

 

160 ... 5000 mbar g

 

See "Mounting flange"

 

5000 mbar g

 

See "Mounting flange"

 

 

 

 

(2.32 ... 72.5 psi g)

 

 

 

(72.5 psi g)

 

 

 

 

Lower measuring limit

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

• Measuring cell with silicone oil filling

-100% of max. span or 30 mbar (0.435 psi a), depending on mounting flange

 

 

Upper measuring limit

 

100% of max. span

 

 

 

100% of the max. nominal measuring range

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Output

 

 

 

 

 

 

 

 

 

 

Output signal

 

4 ... 20 mA

 

 

 

Digital PROFIBUS PA or FOUNDATION Fieldbus

 

 

 

 

 

 

 

 

 

 

 

 

 

 

signal

 

 

 

 

• Lower limit (infinitely adjustable)

3.55 mA, factory preset to 3.84 mA

 

-

 

 

 

 

• Upper limit (infinitely adjustable)

23 mA, factory preset to 20.5 mA or optionally

 

-

 

 

 

 

 

 

set to 22.0 mA

 

 

 

 

 

 

 

 

Load

 

 

 

 

 

 

 

 

 

 

• Without HART communication

RB (UH - 10.5 V)/0.023 A in Ω,

 

 

 

 

 

 

-

 

 

 

 

 

 

UH : Power supply in V

 

 

 

 

 

 

 

 

• With HART communication

RB = 230 ... 500 Ω (SIMATIC PDM) or

 

-

 

 

 

 

 

 

RB = 230 ... 1100 Ω (HART Communicator)

 

 

 

 

 

 

Physical bus

 

-

 

 

 

IEC 61158-2

 

 

 

 

With polarity reversal protection

-

 

 

 

Yes

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Accuracy

 

To EN 60770-1

 

 

 

 

 

 

 

 

Reference conditions

 

Increasing characteristic, start-of-scale value 0 bar, stainless steel seal diaphragm, silicone oil fill-

 

 

(All error data refer always refer to the set span)

ing, room temperature 25 °C (77 °F)) r: Span ratio (r = max. span / set span)

 

 

Error in measurement and fixed-point setting

 

 

 

 

 

 

 

 

 

(including hysteresis and repeatability)

 

 

 

 

 

 

 

 

 

• Linear characteristic

 

 

 

 

 

0.075%

 

 

 

 

 

 

 

 

 

 

 

 

 

- r 10

 

0.15%

 

 

 

 

 

 

 

 

- 10 < r 30

 

0.3%

 

 

 

 

 

 

 

 

- 30 < r 100

 

(0.0075 r + 0.075)%

 

 

 

 

 

 

 

 

Long-term drift (temperature change ±30 °C

(0.25 r)% every 5 years

 

(0.25% every 5 years

 

 

 

 

(±54 °F))

 

static pressure max. 70 bar g (1015 psi g)

 

static pressure max. 70 bar g (1015 psi g)

 

 

Influence of ambient temperature

 

 

 

 

 

 

 

 

 

• at -10 ... +60 °C (14 ... 140 °F)

 

 

 

 

 

 

 

 

 

- 250 mbar (3.63 psi) measuring cell

(0.5 r + 0.2)%

 

 

 

0,7%

 

 

 

 

 

 

(0.4 instead of 0.2 with 10 < r 30)

 

 

 

 

 

 

- 600 mbar (8.7 psi) measuring cell

(0.3 r + 0.2)%

 

 

 

0,5%

 

 

 

 

 

 

(0.4 instead of 0.2 with 10 < r 30)

 

 

 

 

 

 

- 1600 and 5000 mbar (23.2 and 72.5 psi)

(0.25 r + 0.2)% (0.4 instead of 0.2 with 10 < r

 

0,45%

 

 

 

 

measuring cells

 

30)

 

 

 

 

 

 

 

 

• at -40 ... -10 °C and +60

... +85 °C

 

 

 

 

 

 

 

 

 

(-40 ... +14 °F and 140 ...

185 °F)

 

 

 

 

 

 

 

 

 

- 250 mbar (3.63 psi) measuring cell

(0.25 r +0.15)%/10 K

 

 

 

0.4%/10 K

 

 

 

 

 

 

double values with 10 < r 30

 

 

 

 

 

 

- 600 mbar (8.7 psi) measuring cell

(0.15 r +0.15)%/10 K

 

 

 

0.3%/10 K

 

 

 

 

 

 

double values with 10 < r 30

 

 

 

 

 

 

- 1600 and 5000 mbar (23.2 and 72.5 psi)

(0.12 r +0.15)%/10 K

 

 

 

0.27%/10 K

 

 

 

 

measuring cells

 

double values with 10 < r 30

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2/124

Siemens FI 01 · 2009

Page 124
Image 124
Siemens technical specifications 124, Sitrans P, DS III series for level