© Siemens AG 2008

SITRANS P measuring instruments for pressure

Overview

Quick-release clamp-on seals, to DIN 11851 with threaded socket

Quick-release clamp-on seals, with clamp connection

Quick-release clamp-on seals for pressure are available for the following SITRANS P pressure transmitter series:

P300

DS III

DS III PA

DS III FF

Application

The quick-release clamp-on seal is a special design for flowing media and high-viscosity media. Since it is completely inte- grated in the process pipe, no turbulences, dead volumes or other obstructions to the flow occur. The measured medium flows unhindered through the clamp-on seal and results in self- cleaning of the measuring chamber. Furthermore, the clamp-on seal can be cleaned by a pig.

Design

The quick-release clamp is available in two versions:

DIN 11851 with threaded socket

Clamp connection

The clamp-on seal is connected to the pressure transmitter ei- ther directly or by way of a capillary.

Quick-release clamp-on seals

For pressure and absolute pressure

Function

The measured pressure is transferred from the diaphragm,

mounted on the inner circumference of the clamp-on seal, to the filling liquid and then passes through the capillary to the measur- ing chamber of the pressure transmitter. The interior of the

clamp-on seal and of the capillary, as well as the measuring2 chamber of the pressure transmitter, are filled gas-free by the fill-

ing liquid.

Note:

When operating in the low-pressure range, also during commis- sioning, it is recommended to use a vacuum-proof pressure transmitter (see Selection and Ordering data).

Technical specifications

Clamp-on seals of quick-release design for pressure

Connection

Nominal

Nominal

 

diameter

pressure

• To DIN 11851 with threaded

DN 25

PN 40

socket

 

 

 

DN 40

PN 40

 

DN 50

PN 25

 

DN 65

PN 25

 

DN 80

PN 25

 

DN 100

PN 25

• Clamp connection

1½ inch

PN 40

 

2 inch

PN 40

 

2½ inch

PN 40

 

3 inch

PN 40

 

 

 

Material

 

 

• Main body

Stainless steel 1.4404/316L

• Diaphragm

Stainless steel 1.4404/316L

 

 

 

Capillary

 

 

• Length

Max. 10 m (32.8 ft)

• Internal diameter

2 mm (0.079 inch)

• Minimum bending radius

150 mm (5.9 inch)

 

 

Filling liquid

• Food oil (FDA listed)

 

• Glycerin/water (not suitable for

 

use in low-pressure range)

 

 

Permissible ambient temperature

Dependent on the pressure trans-

 

mitter and the filling liquid of the

 

remote seal

 

 

More information can be found in

 

the technical data of the pressure

 

transmitters and in the section

 

"Technical data of filling liquid" in

 

the Technical description to the

 

remote seals

 

 

 

Weight

Approx. 4 kg (approx. 8.82 lb)

 

 

 

Certificate and approvals

 

 

Classification according to pres-

For gases of fluid group 1 and liq-

sure equipment directive

uids of fluid group 1; complies with

(DRGL 97/23/EC)

the requirements of article 3,

 

paragraph 1 (appendix 1);

assigned to category III, confor- mity evaluation module H by the TÜV Nord

Siemens FI 01 · 2009

2/179

Page 179
Image 179
Siemens P technical specifications Quick-release clamp-on seals, For pressure and absolute pressure, 179