Maintenance 5
C A U T I O N
•Do not rotate the rotary switch from position when charging the battery.
•Perform battery charging only with a 9 V
•Disconnect test leads from all the terminals when charging.
•Ensure proper insertion of battery in the multimeter and follow the correct polarity.
Use the specified 24 V DC adaptor to charge the battery. Remember never to turn the rotary switch of the meter when the battery is being charged. Use the following steps to charge the battery:
1Disconnect the test leads from the meter.
2Turn the rotary switch to the position. Connect the power cord to the DC adapter.
3Plug the Red (+)/ Black
adapter to the and the “COM” terminals respectively. The DC adaptor can be replaced with a DC
power supply in order to set the 24 V DC output and the over current limitation to a value more than 0.5 A. Ensure that the polarity of the connection is correct.
4The primary display will show “bAt” and the ‘SbY” will be shown on the secondary display and a short beep will sound to remind you whether you need to charge the battery. Press SHIFT to start charging the battery, or the meter will automatically start the self- test after the 24 V supply is applied. It is recommended that you do not charge if the battery’s capacity is over 90%.
Agilent U1251A/U1252A User’s and Service Guide | 99 |