Performance Tests and Calibration | 7 |
Charge terminal alert test
This test determines if the charge terminal alert is operating properly.
The meter sounds an alert when the ![]()
terminal
detects a voltage level of more than 5 V and the rotary
switch is not set to the ![]()
position. The meter sounds an alert beep and the primary display flashes the “Ch.Err”
until the lead is removed from the
terminal.
N O T E
Figure 6-3 Charge terminal alert
Before conducting this test, make sure the beep function is not disabled at setup.
Agilent U1251A/U1252A User’s and Service Guide | 117 |