Performance Tests and Calibration | 7 |
Charge terminal alert test
This test determines if the charge terminal alert is operating properly.
The meter sounds an alert when the terminal
detects a voltage level of more than 5 V and the rotary
switch is not set to the position. The meter sounds an alert beep and the primary display flashes the “Ch.Err”
until the lead is removed from the
terminal.
N O T E
Figure 6-3 Charge terminal alert
Before conducting this test, make sure the beep function is not disabled at setup.
Agilent U1251A/U1252A User’s and Service Guide | 117 |