7 Performance Tests and Calibration

Step

 

 

Test Function

Range

5520A Output

Error from nominal 1 year

 

 

 

 

 

 

U1251A

U1252A

 

 

 

 

 

 

 

 

 

 

 

 

 

4

Turn the rotary switch to

5 V

5 V

± 2 mV

± 1.75 mV

 

V position (for model U1252A), to

 

 

 

 

 

 

 

V position (for model

 

 

 

 

 

U1251A)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

50 V

50 V

± 20 mV

± 17.5 mV

 

 

 

 

 

 

 

 

 

 

 

 

500 V

500 V

± 200 mV

± 200 mV

 

 

 

 

 

 

 

 

 

 

 

 

1000 V

1000 V

± 800 mV

± 800 mV

 

 

 

 

 

 

 

 

5

Press

button to go to the

5 V

5 V,1 kHz

N/A

± 22.5 mV

 

 

 

V mode [1]

 

5 V, 10 kHz

N/A

± 22.5 mV

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5 V, 20 kHz

N/A

± 41.5 mV

 

 

 

 

 

4.5 V, 100 kHz

N/A

± 169.5 mV

 

 

 

 

 

 

 

 

 

 

 

 

50 V

50 V, 1 kHz

N/A

± 225 mV

 

 

 

 

 

50 V, 10 kHz

N/A

± 225 mV

 

 

 

 

 

50 V, 20 kHz

N/A

± 415 mV

 

 

 

 

 

45 V, 100 kHz

N/A

± 1.695 V

 

 

 

 

 

 

 

 

 

 

 

 

500 V

500 V, 1 kHz

N/A

± 2.25 V

 

 

 

 

 

500 V, 10 kHz

N/A

± 2.25 V

 

 

 

 

 

 

 

 

 

 

 

 

1000 V

1000 V, 1 kHz

N/A

± 8.0 V

 

 

 

 

 

 

 

 

122

Agilent U1251A/U1252A User’s and Service Guide