7 Performance Tests and Calibration
Test Considerations
Long test leads can also act as antennas which may pick up
AC signals.
For optimum performance, all procedures should comply with the following recommendations:
•Ensure that the calibration ambient temperature is stable and is between 18 °C and 28 °C. Ideally the calibration should be performed at 23 °C ± 1 °C.
•Ensure that the ambient relative humidity is less than 80%.
•Allow a warm- up period of five minutes.
•Use shielded twisted pair PTFE- insulated cables to reduce settling and noise errors. Keep the input cables as short as possible.
118 | Agilent U1251A/U1252A User’s and Service Guide |