7 Performance Tests and Calibration

Current terminal test

This test determines if the input warning of the current terminal test is functioning properly.

 

The meter sounds an alert beep when the test lead is

 

inserted into the A terminal but the rotary switch is not set

 

to mA.A function. The primary display will indicate “A- Err”

 

as shown in Figure 6- 2. The primary display will keep flash-

 

ing unless the test lead is removed from “A” terminal.

 

Before conducting this test, make sure the beep function is not disabled in

N O T E

setup.

 

 

 

Figure 6-2Input warning

116