7 Performance Tests and Calibration
Valid adjustment input values
Adjustment can be accomplished using the following input values below.
Table 6-3 Valid adjustment input values
Function | Range | Valid Amplitude Input Values |
|
|
|
| 5 V, 50 V, 500 V, 1000 V | 0.9 to 1.1 x Full Scale |
|
|
|
V | 5 V, 50 V, 500 V, 1000 V | 0.9 to 1.1 x Full Scale |
(for U1251A) |
|
|
|
|
|
| 5 V, 50 V ,500 V, 1000 V | 0.9 to 1.1 x Full Scale |
(for U1252A) |
|
|
|
|
|
| 50 mV, 500 mV, 1000 mV | 0.9 to 1.1 x Full Scale |
|
|
|
| 500 µA, 5000 µA | 0.9 to 1.1 x Full Scale |
|
|
|
| 50 mA, 440 mA, 5 A, 10 A | 0.9 to 1.1 x Full Scale |
|
|
|
Ω | 500 Ω, 5 kΩ, 50 kΩ,500 kΩ, 5 MΩ, | 0.9 to 1.1 x Full Scale |
| 50 MΩ |
|
|
|
|
/ TEMP | 10 nF, 100 nF, 1000 nF, 10 µF, 100 µF, | Make sure to provide 0 °C with ambient |
| 1000 µF, 10 mF / 0 °C | compensation |
|
|
|
134 | Agilent U1251A/U1252A User’s and Service Guide |