Making Measurements 2
If you are working in a constantly varying environment, where ambient temperatures are not constant, do the following:
1Press to select 0 °C compensation. This gives a quick measurement of the relative temperature.
2Avoid contact between the thermocouple probe and the measurement surface.
3After a constant reading is obtained, press to set the reading as the relative reference temperature.
4Touch the measurement surface with the thermocouple probe.
5Read the display for the relative temperature.