Agilent Technologies U1251A manual Fuse checking procedure, Figure above

Models: U1251A

1 185
Download 185 pages 44.17 Kb
Page 125
Image 125

Maintenance 5

Fuse checking procedure

It is recommended that you check the fuses of the multimeter before using it. Follow the instructions below to test the fuses inside the multimeter. Refer to Figure 5- 8 for the respective positions of Fuse 1 and Fuse 2.

1

Set the rotary switch to

.

 

2

Connect the red test lead to the input terminal

.

3To test Fuse 1, place the tip of the test probe on the right

half of input terminal

. Ensure that the probe tip

touches the metal inside the input terminal, as shown in the

figure above.

Agilent U1251A/U1252A User’s and Service Guide

105

Page 125
Image 125
Agilent Technologies U1251A manual Fuse checking procedure, Figure above