1 ms Peak Hold

The Peak Hold function allows the measurement of peak

voltage for analysis of components such as power

distribution transformers and power factor correction

capacitors. The peak voltage obtained can be used to deter-

mine the crest factor:

Crest factor = Peak value/True RMS value

1Press for more than 1 second to toggle 1 ms Peak Hold mode ON / OFF.

2Press to scroll through maximum and minimum

peak readings. MAX indicates the maximum peak,

while MIN indicates the minimum peak.

N O T E

If the reading is "OL", press

to change the measuring range and

to re-start the peak-recording measurement.

 

If you need to re-start the peak recording, press .

3Press and hold or to exit this mode.

for more than 1 second

4According to the measurements shown in Figure 3- 6 on page 60, the crest Factor will be 2.5048/1.768 =1.416.