7

Performance Tests and Calibration

 

 

 

 

 

 

 

 

 

 

 

 

 

Step

Test Function

Range

5520A Output

Error from nominal 1 year

 

 

 

 

U1251A

U1252A

 

 

 

 

 

 

 

 

 

 

 

 

 

 

8

Turn the rotary switch to the W

500 W

500 W

± 500 mW [3]

± 350 mW [3]

 

 

position

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5 kW

5 kW

± 4.5 W [3]

± 3 W [3]

 

 

 

 

50 kW

50 kW

± 45 W

± 30 W

 

 

 

 

 

 

 

 

 

 

 

 

500 kW

500 kW

± 450 W

± 300 W

 

 

 

 

 

 

 

 

 

 

 

 

5 MW

5 MW

± 10.5 kW

± 8 kW

 

 

 

 

 

 

 

 

 

 

 

 

50 MW [4]

50 MW

± 0.510 MW

± 0.505 MW

 

 

 

 

500 MW

450 MW

N/A

± 36.1 MW

 

 

 

 

 

 

 

 

 

9

Press

button to go to ns

500 nS [5]

50 nS

± 0.7 nS

± 0.6 nS

 

 

mode

 

 

 

 

 

 

 

 

 

 

 

 

 

10

Turn the rotary switch to the Hz/

Diode

1 V

± 1 mV

± 1 mV

 

 

position (for model

 

 

 

 

 

 

U1252A), to

position (for

 

 

 

 

 

 

model U1251A)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

33250A Output

 

 

 

 

 

 

 

 

 

 

 

11

Press

button to go to

999.99 kHz

200 mVrms,

N/A

± 52 Hz

 

 

frequency counter mode [6]

 

100 kHz

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

12

Press

button to go to

99.999 MHz

600 mVrms,

N/A

± 5.2 kHz

 

 

frequency counter mode divide

 

10 MHz

 

 

 

 

by 100

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5520A Output

 

 

 

 

 

 

 

 

 

 

13

Turn the rotary switch to the

10.000 nF

10.000 nF

± 0.108 nF

± 0.108 nF

 

 

TEMP/

position [7]

 

 

 

 

 

 

 

 

100.00 nF

100.00 nF

± 1.05 nF

± 1.05 nF

 

 

 

 

 

 

 

 

 

 

 

 

1000.0 nF

1000.0 nF

± 10.5 nF

± 10.5 nF

 

 

 

 

 

 

 

 

 

 

 

 

10.000 μF

10.000 μF

± 0.105 μF

± 0.105 μF

 

 

 

 

 

 

 

 

124

Agilent U1251A/U1252A User’s and Service Guide

Page 144
Image 144
Agilent Technologies U1251A manual ± 350 mW, 33250A Output